Thrombin-Accelerated Quick Clotting Serum Tubes: An Evaluation with 22 Common Biochemical Analytes
Authors:
Open Access
- 3 April 2013
- journal article
- research article
- Published by Wiley in Advances in Hematology
- Vol. 2013, 1-8
Abstract
Clot activator serum tubes have significantly improved turnaround times for result reporting compared to plain tubes. With increasing workload and service performance expectations confronting clinical laboratories with high-volume testing and with particular emphasis on critical analytes, attention has focussed on preanalytical variables that can be improved. We carried out a field study on the test performance of BD vacutainer rapid serum tubes (RSTs) compared to current institutional issued BD vacutainer serum separator tubes (SSTs) in its test result comparability, clotting time, and stability on serum storage. Data from the study population () of patients attending outpatient clinics and healthy subjects showed that results for renal, liver, lipids, cardiac, thyroid, and prostate biochemical markers were comparable between RSTs and SSTs. Clotting times of the RSTs were verified to be quick with a median time of 2.05 min. Analyte stability on serum storage at 4°C showed no statistically significant deterioration except for bicarbonate, electrolytes, and albumin over a period of 4 days. In conclusion, RSTs offered savings in the time required for the clotting process of serum specimens. This should translate to further trimming of the whole process from blood collection to result reporting without too much sacrifice on test accuracy and performance compared to the current widely used SSTs in most clinical laboratories.
Keywords
This publication has 14 references indexed in Scilit:
- BD rapid serum tubes reduce false positive plasma troponin T results on the Roche Cobas e411 analyzerClinical Biochemistry, 2012
- Use of the BD Vacutainer Rapid Serum Tube Reduces False-Positive Results for Selected Beckman Coulter Unicel DxI ImmunoassaysAmerican Journal of Clinical Pathology, 2011
- Evaluation of the Becton-Dickinson rapid serum tube: does it provide a suitable alternative to lithium heparin plasma tubes?Clinical Chemistry and Laboratory Medicine (CCLM), 2010
- Improving laboratory efficiencies through significant time reduction in the preanalytical phaseClinical Chemistry and Laboratory Medicine (CCLM), 2009
- Analysis of turnaround time by subdividing three phases for outpatient chemistry specimens.2009
- Selected Analyte Values in Serum Versus Heparinized Plasma Using the SYNCHRON LX PRO Assay Methods/InstrumentLaboratory Medicine, 2006
- Comparison of Serum and Heparinized Plasma Samples for Measurement of Chemistry AnalytesClinical Chemistry, 2004
- Plasma or serum samples: measurements of cardiac troponin T and of other analytes comparedClinical Chemistry and Laboratory Medicine (CCLM), 2004
- On the Comparison of Serum and Plasma Samples in Troponin AssaysClinical Chemistry, 2003
- Addition of 3-Deazaadenosine to Vacutainer Tubes Stabilizes Whole-Blood Homocysteine for At Least 6 Hours at Ambient TemperatureClinical Chemistry, 2003