Protein sorption on polymer surfaces measured by fluorescence labels
Authors:
- 31 December 1977
- journal article
- research article
- Published by Wiley in Journal of Biomedical Materials Research
- Vol. 12 (1) , 55-65
Abstract
Fluorescence labeling can be used in studying protein sorption on various surfaces with a sensitivity of about 10−8 g/cm2, commensurate with radioactive labeling. Fluorescamine proved to be the most suitable compound for studying protein sorption on hydrophilic gels, because, unlike fluoresceine isothiocyanate and dansylchloride, free fluorochrome does not interfere with measurements. Sorption properties of labeled serum albumin were tested on poly(2‐hydroxyethyl methacrylate), on the copolymer of 2‐hydroxyethyl methacrylate with methyl methacrylate, and on polyethylene. Labeling does not cause aggregation of the protein, but, as expected, it shifts and somewhat broadens its electrophoretic band while at the same time slightly raising its affinity toward hydrophobic surfaces.Keywords
This publication has 7 references indexed in Scilit:
- Fibrinogen, Globulins, Albumin and Plasma at Interfaces1975
- Multiple internal reflection fluorescence spectrometryAnalytical Chemistry, 1973
- Fluorometric assay of proteins in the nanogram rangeArchives of Biochemistry and Biophysics, 1973
- WATER AS A BIOMATERIALASAIO Journal, 1973
- Novel reagent for the fluorometric assay of primary aminesJournal of the American Chemical Society, 1972
- Initial events in interactions of blood with a foreign surfaceJournal of Biomedical Materials Research, 1969
- Polarization of the fluorescence of macromolecules. 2. Fluorescent conjugates of ovalbumin and bovine serum albuminBiochemical Journal, 1952