A Reliable Semiautomated Method for the Determination of Total Thiamine in Whole Blood by the Thiochrome Method with High-Performance Liquid Chromatography

Authors:

Abstract
A sensitive and reliable method for the determination of total thiamine (vitamin B 1 ) in whole blood has been developed which is suited for routine analysis. After extraction, and enzymatic hydrolysis of thiamine phosphate esters, thiamine is separated from interfering compounds by a fully automated high-performance liquid chromatographic (HPLC) system. Thiamine is detected fluorometrically as thiochrome. By calculating the concentration of thiamine on-line with the aid of a computer, it is possible to complete one analysis within four hours. Routine thiamine determinations can be carried out in a series of 120 samples within 48 hours. The within-assay and between-assay coefficients of variation of the analysis of total thiamine in whole blood were 4·2 and 4·4%, respectively. The between-assay analytical recovery of thiamine diphosphate added to blood samples was 99·9 ± 11·7% (mean ± SD). The HPLC method described has been applied also to the analysis of thiamine in plasma and erythrocytes. In agreement with other reports, it was found that about 80% of total thiamine of whole blood is present in the erythrocytes. Reference values of thiamine in whole blood of the human were found in the range 95–155 nmol/l.