Spontaneous magnetoencephalography alpha rhythm measurement in a cylindrical magnetic shield employing magnetic shaking
Authors:
- 9 May 2003
- journal article
- conference paper
- Published by AIP Publishing in Journal of Applied Physics
- Vol. 93 (10) , 6733-6735
Abstract
This article shows a demonstration of magnetoencephalography measurement in a cylindrical magnetic shield made of cobalt-based amorphous tape with magnetic shaking. The noise levels of the first-order superconducting quantum interference device gradiometers that operated in the shield were reduced to as low as at 10 Hz by surrounding it with a simple rf shield made of conductive cloth. We observed spontaneous alpha rhythms from a human brain in this shielding system. Alpha rhythms and their suppression caused by opening the eyes were clearly found, which was also confirmed by electroencephalography measurement from the same volunteer under similar conditions.
Keywords
This publication has 6 references indexed in Scilit:
- Measurement of RF Noise in a Magnetic Shield with a Developed Monopole Antenna. Effect of RF Noise on FLL Operation of dc-SQUID Sensors.Journal of the Magnetics Society of Japan, 2002
- Design of a large-scale vertical open-structure cylindrical shield employing magnetic shakingJournal of Applied Physics, 2000
- Effect of RF interference on characteristics of DC SQUID systemIEEE Transactions on Applied Superconductivity, 1993
- Effective shielding for low-level magnetic fieldsJournal of Applied Physics, 1988
- Biomagnetic instrumentationReview of Scientific Instruments, 1982
- Magnetoencephalography: Evidence of Magnetic Fields Produced by Alpha-Rhythm CurrentsScience, 1968