Electromagnetic interference of implantable cardiac devices from a shoulder massage machine
- 8 April 2014
- journal article
- Published by Springer Nature in Journal of Artificial Organs
- Vol. 17 (3) , 243-249
Abstract
Shoulder massage machines have two pads that are driven by solenoid coils to perform a per cussive massage on the shoulders. There have been concerns that such machines might create electromagnetic interference (EMI) in implantable cardiac devices because of the time-varying magnetic fields produced by the alternating current in the solenoid coils. The objective of this study was to investigate the potential EMI from one such shoulder massage machine on implantable cardiac devices. We measured the distribution profile of the magnetic field intensity around the massage machine. Furthermore, we performed an inhibition test and an asynchronous test on an implantable cardiac pacemaker using the standardized Irnich human body model. We examined the events on an implantable cardioverter–defibrillator (ICD) using a pacemaker programmer while the massage machine was in operation. The magnetic field distribution profile exhibited a peak intensity of 212 (A/m) in one of the solenoid coils. The maximal interference distance between the massage machine and the implantable cardiac pacemaker was 28 cm. Ventricular fibrillation was induced when the massage machine was brought near the electrode of the ICD and touched the Irnich human body model. It is necessary to provide a “don’t use” warning on the box or the exterior of the massage machines or in the user manuals and to caution patients with implanted pacemakers about the dangers and appropriate usage of massage machines.Keywords
This publication has 6 references indexed in Scilit:
- Electronic Article Surveillance Systems and Interactions With Implantable Cardiac Devices: Risk of Adverse Interactions in Public and Commercial SpacesMayo Clinic Proceedings, 2007
- Electronic Article Surveillance Systems and Interactions With Implantable Cardiac Devices: Risk of Adverse Interactions in Public and Commercial SpacesMayo Clinic Proceedings, 2007
- Paradoxical Atrial Undersensing: Noise Rate Reversion or Amplifier Ringing?Journal of Cardiovascular Electrophysiology, 2006
- Electromagnetic Interference of Implantable Unipolar Cardiac Pacemakers by an Induction OvenPacing and Clinical Electrophysiology, 2005
- Electromagnetic Interference of Pacemakers by Mobile PhonesPacing and Clinical Electrophysiology, 1996