2D model‐based reconstruction for magnetic particle imaging
Authors:
- Tobias KnoppTKTobias Knopp
- Sven BiedererSBSven Biederer
- Timo F. SattelTSTimo F. Sattel
- Jürgen RahmerJRJürgen Rahmer
- Jürgen WeizeneckerJWJürgen Weizenecker
- Bernhard GleichBGBernhard Gleich
- Jörn BorgertJBJörn Borgert
- Thorsten M. BuzugTBThorsten M. Buzug
- 7 January 2010
- journal article
- Published by Wiley in Medical Physics
- Vol. 37 (2) , 485-491
Abstract
Purpose: Magnetic particle imaging (MPI) is a new quantitative imaging technique capable of determining the spatial distribution of superparamagnetic nanoparticles at high temporal and spatial resolution. For reconstructing this spatial distribution, the particle dynamics and the scanner properties have to be known. To date, they are obtained in a tedious calibration procedure by measuring the magnetization response of a small delta sample shifted through the measuring field. Recently, first reconstruction results using a 1D model‐based system function were published, showing comparable image quality as obtained with a measured system function. In this work, first 2D model‐based reconstruction results of measured MPI data are presented.Methods: To simulate the system function, various parameters have to be modeled, namely, the magnetic field, the particle magnetization, the voltage induced in the receive coils, and the transfer function of the receive chain. To study the accuracy of the model‐based approach, 2D MPI data are measured and reconstructed with modeled and measured system functions.Results: It is found that the model‐based system function is sufficiently accurate to allow for reconstructing experimental data. The resulting image quality is close to that obtained with a measurement‐based reconstruction.Conclusions: The model‐based system function approach addresses a major drawback of the measurement‐based procedure, namely, the long acquisition time. In this work, the acquisition of the measurement‐based system function took, while the model‐based system function was obtained in only . For 3D data, where the acquisition of the measurement‐based system function takes more than , the need for an efficient system function generation is even more obvious.Keywords
This publication has 10 references indexed in Scilit:
- Magnetization response spectroscopy of superparamagnetic nanoparticles for magnetic particle imagingJournal of Physics D: Applied Physics, 2009
- Model-Based Reconstruction for Magnetic Particle ImagingIEEE Transactions on Medical Imaging, 2009
- Signal encoding in magnetic particle imaging: properties of the system functionBMC Medical Imaging, 2009
- Three-dimensional real-timein vivomagnetic particle imagingPhysics in Medicine & Biology, 2009
- Trajectory analysis for magnetic particle imagingPhysics in Medicine & Biology, 2008
- Frequency distribution of the nanoparticle magnetization in the presence of a static as well as a harmonic magnetic fieldMedical Physics, 2008
- Experimental results on fast 2D-encoded magnetic particle imagingPhysics in Medicine & Biology, 2008
- A simulation study on the resolution and sensitivity of magnetic particle imagingPhysics in Medicine & Biology, 2007
- Tomographic imaging using the nonlinear response of magnetic particlesNature, 2005
- Magnetic Iron Oxide Particles Coated with Carboxydextran for Parenteral Administration and Liver ContrastingActa Radiologica, 1997