Dose Exposure of Patients Undergoing Comprehensive Stroke Imaging by Multidetector-Row CT: Comparison of 320-Detector Row and 64-Detector Row CT Scanners
Open Access
- 28 January 2010
- journal article
- Published by American Society of Neuroradiology (ASNR) in American Journal of Neuroradiology
- Vol. 31 (6) , 1003-1009
Abstract
BACKGROUND AND PURPOSE: Recently introduced 320-detector row CT enables whole brain perfusion imaging compared to a limited scanning area in 64-detector row CT. Our aim was to evaluate patient radiation exposure in comprehensive stroke imaging by using multidetector row CT consisting of standard CT of the head, CTA of cerebral and cervical vessels, and CTP. MATERIAL AND METHODS: Organ doses were measured by using LiF-TLDs located at several organ sites in an Alderson-Rando phantom. Effective doses were derived from these measurements. Stroke protocols including noncontrast head CT, CTA of cerebral and cervical vessels, and CTP were performed on 320- and 64-detector row scanners. RESULTS: Measured effective doses for the different scanning protocols ranged between 1.61 and 4.56 mSv, resulting in an effective dose for complete stroke imaging of 7.52/7.54 mSv (m/f) for 64-detector row CT and 10.56/10.6 mSv (m/f) for 320-detector row CT. The highest organ doses within the area of the primary beam were measured in the skin (92 mGy) and cerebral hemispheres (69.91 mGy). Use of an eye-protection device resulted in a 54% decrease of the lens dose measured for the combo protocol for whole-brain perfusion with the 320-detector row CT scanner. CONCLUSIONS: Phantom measurements indicate that comprehensive stroke imaging with multidetector row CT may result in effective radiation doses from 7.52 mSv (64-detector row CT) to 10.6 mSv (320-detector row CT). The technique of 320-detector row CT offers additional information on the time course of vascular enhancement and whole-brain perfusion. Physicians should weigh the potential of the new technique against the higher radiation dose that is needed. Critical doses that would cause organ damage were not reached.Keywords
This publication has 17 references indexed in Scilit:
- Computed tomography dose assessment for a 160 mm wide, 320 detector row, cone beam CT scannerPhysics in Medicine & Biology, 2009
- Concepts for dose determination in flat-detector CTPhysics in Medicine & Biology, 2008
- The trouble withMedical Physics, 2007
- Conversion factor for CT dosimetry to assess patient dose using a 256-slice CT scannerBritish Journal of Radiology, 2006
- Comparison of patient doses in 256-slice CT and 16-slice CT scannersBritish Journal of Radiology, 2006
- Enlarged longitudinal dose profiles in cone‐beam CT and the need for modified dosimetryMedical Physics, 2005
- Cerebral Perfusion CT: Technique and Clinical ApplicationsRadiology, 2004
- The measurement of radiation dose profiles for electron-beam computed tomography using film dosimetryMedical Physics, 1994
- Radiation dosimetry survey of computed tomography systems from ten manufacturersBritish Journal of Radiology, 1982
- A Film Dosimetry System for Use in Computed TomographyRadiology, 1978