Low-Cost, Smartphone-Based Specular Imaging and Automated Analysis of the Corneal Endothelium
Authors:
Open Access
- 1 April 2021
- journal article
- research article
- Published by Association for Research in Vision and Ophthalmology (ARVO) in Translational Vision Science & Technology
- Vol. 10 (4) , 4
Abstract
Purpose: Specular and confocal microscopes are important tools to monitor the health of the corneal endothelium (CE), but their high costs significantly limit accessibility in low-resource environments. We developed and validated a low-cost, fully automated method to quantitatively evaluate the CE using smartphone-based specular microscopy. Methods: A OnePlus 7 Pro smartphone attached to a Topcon SL-D701 slit-lamp was used to image the central corneal endothelium of 30 eyes using the specular reflection technique. A novel on-device image processing algorithm automatically computed endothelial cell density (ECD), percentage of hexagonal cells (HEX), and coefficient of variation (CV) values. These values were compared with the ECD, HEX, and CV generated by a Torrey EM-4000 specular microscope used to image the same set of eyes. Results: No significant differences were found in ECD (2799 +/- 156 cells/mm(2) vs. 2779 +/- 166 cells/mm(2) ; P = 0.28) and HEX (52 +/- 6% vs. 53 +/- 6%; P = 0.50) computed by smartphone-based specular imaging and specular microscope, respectively. A statistically significant difference in CV (34 +/- 3% vs. 30 +/- 3%; P < 0.01) was found between the two methods. The concordance achieved between the smartphone-based method and the Torrey specular microscope is very similar to the concordance between two specular microscopes reported in the literature. Conclusions: Smartphone-based specular imaging and automated analysis is a lowcost method to quantitatively evaluate the CE with accuracy comparable to the clinical standard. Translational Relevance: This tool can be used to screen the CE in low-resource regions and prompt investigation of suspected corneal endotheliopathies.Keywords
All Related Versions
This publication has 16 references indexed in Scilit:
- Smartphone-based imaging of the corneal endothelium at sub-cellular resolutionJournal of Modern Optics, 2016
- Influence of applied corneal endothelium image segmentation techniques on the clinical parametersComputerized Medical Imaging and Graphics, 2016
- Report of the Eye Bank Association of America Medical Review Subcommittee on Adverse Reactions Reported From 2007 to 2014Cornea, 2016
- The ownership and clinical use of smartphones by doctors and nurses in the UK: a multicentre survey studyBMJ Innovations, 2015
- Comparison of corneal endothelial cell measurements by two non-contact specular microscopesBMC Ophthalmology, 2015
- Mobilizing slit lamp to the field: A new affordable solutionIndian Journal Of Ophthalmology, 2015
- Prevalence of Corneal Dystrophies in the United States: Estimates from Claims DataInvestigative Ophthalmology & Visual Science, 2011
- Evaluation of the Corneal Endothelium Using Noncontact and Contact Specular MicroscopyCornea, 2011
- Biology of the corneal endothelium in health and diseaseEye, 2003
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986