Verification of the intraluminal multiple electrical impedance measurement for the recording of gastrointestinal motility
Authors:
- 28 June 1993
- journal article
- Published by Wiley in Neurogastroenterology & Motility
- Vol. 5 (2) , 107-122
Abstract
A new intraluminal electrical impedance procedure for high‐resolution measurements and the quantitative assessment of gastrointestinal motility is examined in healthy volunteers by cineradiography and manometry. The peristalsis in the oesophagus, stomach and small bowel is recorded with a combined impedance‐pressure catheter. Additionally, investigations with a flexible 16‐channel impedance catheter with a closed surface and a diameter of 3 mm are carried out in the oesophagus and small intestine. The 16 measuring segments with a length of 2 cm each record the contractile patterns from a 32‐cm‐long organ section without a gap. The correctness of the physical approach is validated by concurrent impedance and cineradiography recordings of the oesophageal peristalsis. The comparative studies confirm a close relation between the pressure and the impedance changes in the oesophagus and small intestine. The time analysis of the impedance tracings offers information about the bolus transit and the change of the wall compliance along the organ. Regions of high or low compliance of the muscular wall can be recognized. From the impedance tracings the direction of the contraction waves as well as their velocities, lengths, beginnings and ends can be determined. Moreover, the beginning and end of the bolus, and from them the momentary bolus length, can be characterized for any instant during the bolus transit.Keywords
This publication has 6 references indexed in Scilit:
- Intraluminal Multiple Electric Impedance Procedure for Measurement of Gastrointestinal MotilityNeurogastroenterology & Motility, 1991
- A time-dependent study of passive esophageal wall properties and collagen content in rabbits with esophageal varicesDigestive Diseases and Sciences, 1991
- The four-electrode impedance technique: a method for investigation of compliance in luminal organsClinical Physics and Physiological Measurement, 1988
- Improved Infusion System for Intraluminal Esophageal ManometryGastroenterology, 1977
- Factors Determining Pressure Measurement Accuracy by Intraluminal Esophageal ManometryGastroenterology, 1976
- The specific resistance of biological material—A compendium of data for the biomedical engineer and physiologistMedical & Biological Engineering & Computing, 1967