Pacemaker Tachycardia in a Minute Ventilation Rate‐Adaptive Pacemaker Induced by Electrocardiographic Monitoring
- 6 April 2006
- journal article
- case report
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 29 (4) , 438-440
Abstract
Programmed upper rate pacing occurred in a patient with a rate‐adaptive pacemaker when he was connected to a cardiac monitor in an emergency department. The tachycardia was mistakenly interpreted to be ventricular tachycardia and the patient received multiple DC shocks as well as intravenous amiodarone and sotalol, resulting in severe hemodynamic deterioration. It is important that physicians working in the hospital environment be familiar with this pacemaker‐monitor interaction as the problem may be easily rectified by disconnecting the monitor or by reprogramming the pacemaker to a nonrate‐adaptive pacing mode.Keywords
This publication has 5 references indexed in Scilit:
- Monitoring equipment induced tachycardia in patients with minute ventilation rate-responsive pacemakersBritish Journal of Anaesthesia, 2000
- Supraventricular Tachycardia Induced by Datex Patient Monitoring SystemAnesthesia & Analgesia, 1998
- Inappropriate Rate Change in Minute Ventilation Rate Responsive Pacemakers Due to Interference by Cardiac MonitorsPacing and Clinical Electrophysiology, 1997
- Upper Limit Ventricular Stimulation in Respiratory Rate Responsive Pacing due to ElectrocauteryPacing and Clinical Electrophysiology, 1989