Measuring Soil and Tissue Potassium with a Portable Ion-Specific Electrode in Cotton
Authors:
- 10 October 2016
- journal article
- research article
- Published by Taylor & Francis in Communications in Soil Science and Plant Analysis
- Vol. 47 (18) , 2148-2155
Abstract
Most laboratories make potassium (K) fertilizer recommendations based on field research calibrated with soil K from atomic absorption spectroscopy, flame emission spectroscopy, or inductively coupled plasma mass spectrometry. Information is needed to interpret readings from K+ meters on cotton. The objective was to compare soil and cotton plant sap potassium content from a Laqua Twin™ K+ meter to results from standard tests. Aluminum sulfate solution was used with the meter to extract K from 10 soil samples from the North America Proficiency Testing program. A linear relationship (R2 = 0.86) was found between K measured with a K+ meter and K measured with 138 labs using ammonium acetate or Mehlich-3. Using a factor of 1.4754 to adjust readings, fertilizer recommendations with the K+ meter were similar to results from atomic absorption spectroscopy. A linear relationship was found between petiole K measured in the lab and petiole sap measured with the K+ meter.Keywords
Funding Information
- Cotton (Project 13573MO)
This publication has 3 references indexed in Scilit:
- Managing Potassium and Phosphorus Soil Test Levels in a Rice–Soybean RotationCrop, Forage & Turfgrass Management, 2015
- Accuracy Varies for Commercially Available Soil Test Kits Analyzing Nitrate–Nitrogen, Phosphorus, Potassium, and pHHortTechnology, 2007
- Measuring K+ in Rice Basal Stem Sap with a Cardy MeterCrop Management, 2004