A Gaze-Guided Human Locomotion Intention Recognition and Volitional Control Method for Knee-Ankle Prostheses
- 16 April 2025
- journal article
- Published by Institute of Electrical and Electronics Engineers (IEEE) in IEEE Transactions on Cognitive and Developmental Systems
- Vol. 18 (1) , 22-32
Abstract
This paper presents a novel gaze-guided volitional control method for knee-ankle prostheses, designed to enhance the precision and intuitiveness of prosthetic control in complex locomotion tasks. The method utilizes an eye-tracking system coupled with an RGBD camera to estimate 3D gaze points, which are then used to construct a gaze attention map. This map serves as the basis for predicting human locomotion intentions, enabling the prosthesis to accurately step on the intended foot placement. Experimental validation was conducted with both a healthy subject and an amputee subject, focusing on level-ground walking and stair ascent tasks. The results demonstrated that the prosthesis could achieve precise foot placement, with a mean error of 1.22 cm ± 3.14 cm for the healthy subject and 3.40 cm ± 0.66 cm for the amputee subject during level-ground walking. In stair ascent tasks, the system successfully adapted to various gait patterns, with the prosthetic joint angles closely matching natural human movements. The proposed method also effectively reduced the root mean square error (RMSE) of 3D gaze estimation to 2.70 cm, 2.63 cm, and 1.97 cm along the x, y, and z axes, respectively. These findings indicate that the gaze-guided volitional control method not only enhances the functionality of knee-ankle prostheses but also offers a more intuitive interface for users, aligning prosthetic actions with the user’s visual focus and intentions.Keywords
Funding Information
- National Natural Science Foundation of China (U1913205)
- National Natural Science Foundation of China (62103180)
- Basic and Applied Basic Research Foundation of Guangdong Province (2024B1515020008)
- Science, Technology and Innovation Commission of Shenzhen Municipality (KCXFZ20230731093401004)
- Southern University of Science and Technology, Shenzhen, China (G03034K003)
- Special Funds for the Cultivation of Guangdong College Students’ Scientific and Technological Innovation (pdjh2025c10802)
This publication has 31 references indexed in Scilit:
- Enhancing Prosthetic Safety and Environmental Adaptability: A Visual-Inertial Prosthesis Motion Estimation Approach on Uneven Terrains2024
- Volitional EMG Control Enables Stair Climbing with a Robotic Powered Knee Prosthesis2013 IEEE/RSJ International Conference on Intelligent Robots and Systems, 2023
- Natural Grasp Intention Recognition Based on Gaze in Human–Robot InteractionIEEE Journal of Biomedical and Health Informatics, 2023
- Motion Intention Prediction and Joint Trajectories Generation Toward Lower Limb Prostheses Using EMG and IMU SignalsIEEE Sensors Journal, 2022
- A comprehensive, open-source dataset of lower limb biomechanics in multiple conditions of stairs, ramps, and level-ground ambulation and transitionsJournal of Biomechanics, 2021
- Stepping in the direction of the fall: the next foot placement can be predicted from current upper body state in steady-state walkingBiology Letters, 2014
- Robust Statistics2011
- A flexible new technique for camera calibrationIEEE Transactions on Pattern Analysis and Machine Intelligence, 2000
- Understanding the roles of vision in the control of human locomotionGait & Posture, 1997
- Decay Theory of Immediate MemoryNature, 1957