Abstract
A pressure-sensing system with a half-Wheatstone-bridge-to-digital converter (HBDC) is proposed for implantable intraocular pressure (IOP) monitoring systems. The half-Wheatstone-bridge (H-WhB) sensor uses an RC-delay comparison instead of direct bias of the pressure transducer, allowing it to self-limit the current for efficient operation. To overcome the limited sensitivity of the H-WhB, energy-efficient bit-level oversampling (OS) is introduced. The system achieves 0.24-mmHg (1 $\sigma )$ resolution with an 8.58-nJ $\cdot$ mmHg $^{2}$ FOM and 12.79- $\mu $ W power consumption with a 11.52-ms conversion time. This marks a 2.8 $\times$ improvement in measured system power, 56.24 $\times$ improvement in energy consumption, and 2.8 $\times$ improvement in resolution compared to the prior H-WhB-based pressure sensor. In addition, the HBDC overcomes the low sensitivity limitations of the H-WhB and achieves a resolution FOM comparable to that of Wheatstone-bridge (WhB) sensor-based pressure sensors.