Detection of bacteria in burn wounds with a novel handheld autofluorescence wound imaging device: a pilot study
Authors:
- 2 August 2019
- journal article
- research article
- Published by Mark Allen Group in Journal of Wound Care
- Vol. 28 (8) , 548-554
Abstract
Objective: To compare the detection of bacteria in burn wounds between an bacterial fluorescence imaging device MolecuLight i:X, (Canada), and standard microbiological swabs. Methods: Wounds were swabbed three times on one occasion; once with a standard swab, once with a high-fluorescent area swab, indicating a bacterial load >104 colony-forming units (CFU)/gram and a finally with a non-fluorescent (nF) area swab. Proportion agreement of the microbiological results was calculated and the accuracy of the device to detect relevant bacteria was assessed. Results: A total of 14 patients with 20 wounds participated in the study. Median post-burn day at sampling time was 21 days. Of the 20 wounds, nine had a positive swab result in either of the three swabs, and 11 showed a highfluorescent area. Overall, positive and negative proportion agreement between standard swab and high-fluorescent swab sample results were 100%. Sensitivity, specificity, positive and negative predictive values of presence of high-fluorescence were 78%, 64%, 64%, and 78%, respectively. For Pseudomonas aeruginosa detection, these results were 100%, 70%, 44% and 100%, respectively. Conclusion: The diagnostic accuracy of the bacterial fluorescence imaging device to detect relevant bacteria in burn wounds was moderate and the reliability was equal to standard swabbing. Further research in larger sample sizes and on the relevance of minimal bacterial load and its potential to help with Pseudomonas aeruginosa management is needed.Keywords
This publication has 15 references indexed in Scilit:
- Improved detection of clinically relevant wound bacteria using autofluorescence image‐guided sampling in diabetic foot ulcersInternational Wound Journal, 2017
- Point-of-Care Autofluorescence Imaging for Real-Time Sampling and Treatment Guidance of Bioburden in Chronic Wounds: First-in-Human ResultsPLOS ONE, 2015
- Common pathogens in burn wound and changes in their drug sensitivityBurns, 2011
- Prophylactic antibiotics for burns patients: systematic review and meta-analysisBMJ, 2010
- Changes in bacterial isolates from burn wounds and their antibiograms: A 20-year study (1986–2005)Burns, 2009
- Burn Wound InfectionsClinical Microbiology Reviews, 2006
- The Risk Factors and Time Course of Sepsis and Organ Dysfunction after Burn Trauma2003
- QUANTITATIVE OPTICAL SPECTROSCOPY FOR TISSUE DIAGNOSISAnnual Review of Physical Chemistry, 1996
- Porphyrin Photosensitization of Bacteria1985
- Predicting Fatal Sepsis in Burn Patients1979