Digital radiology using active matrix readout of amorphous selenium: Construction and evaluation of a prototype real‐time detector
Authors:
- Wei ZhaoWZWei Zhao
- Ira BlevisIBIra Blevis
- Stephen GermannSGStephen Germann
- J. A. RowlandsJRJ. A. Rowlands
- David WaechterDWDavid Waechter
- Zhongshou HuangZHZhongshou Huang
- 1 December 1997
- journal article
- Published by Wiley in Medical Physics
- Vol. 24 (12) , 1834-1843
Abstract
The goal of the present work is to develop a large area, flat‐panel solid‐state detector for both digital radiography and fluoroscopy. The proposed detector employs a photoconductive layer of amorphous selenium (‐Se) to convert x rays into charge. The charge image formed by the ‐Se layer is electronically read out in situ using a two dimensional array of thin film transistors (TFTs), or active matrix. Since the active matrix readout is capable of producing x‐ray images in real‐time, it can potentially be applied in both radiography and fluoroscopy. In this paper, the imaging performance of this concept is investigated using a prototype x‐ray imaging detector. The designs for the active matrix, the peripheral electronic circuits, and the image acquisition system are described. Measurements of x‐ray imaging properties of the prototype detector, i.e., x‐ray sensitivity, presampling modulation transfer function (MTF), and noise power spectrum (NPS), were performed, and from which the spatial frequency dependent detective quantum efficiency (DQE) of the prototype was derived. The experimental results are in agreement with the results of our theoretical analysis. The factors affecting the imaging performance and methods of improvement in the future are discussed.Keywords
This publication has 17 references indexed in Scilit:
- Digital radiology using active matrix readout of amorphous selenium: Theoretical analysis of detective quantum efficiencyMedical Physics, 1997
- X‐ray imaging using amorphous selenium: Feasibility of a flat panel self‐scanned detector for digital radiologyMedical Physics, 1995
- X‐ray imaging using amorphous selenium: Inherent spatial resolutionMedical Physics, 1995
- Effects of undersampling on the proper interpretation of modulation transfer function, noise power spectra, and noise equivalent quanta of digital imaging systemsMedical Physics, 1995
- Fluoroscopic x-ray imaging with amorphous silicon thin-film arrays1994
- Demonstration of megavoltage and diagnostic x‐ray imaging with hydrogenated amorphous silicon arraysMedical Physics, 1992
- Thin film cadmium selenide technology in large area active matrix high resolution displaysMicroelectronic Engineering, 1992
- X-ray imaging using amorphous selenium: Determination of x-ray sensitivity by pulse height spectroscopyMedical Physics, 1992
- Investigation of basic imaging properties in digital radiography. 2. Noise Wiener spectrumMedical Physics, 1984
- The line spread function and modulation transfer function of a computed tomographic scannerMedical Physics, 1976