The effect of ambient pressure on well chamber response: Monte Carlo calculated results for the HDR 1000 Plus
Authors:
- Tim D. BohmTBTim D. Bohm
- Sheridan L. GriffinSGSheridan L. Griffin
- Paul M. DeLucaPDPaul M. DeLuca
- Larry A. DeWerdLDLarry A. DeWerd
- 25 March 2005
- journal article
- Published by Wiley in Medical Physics
- Vol. 32 (4) , 1103-1114
Abstract
The determination of the air kerma strength of a brachytherapy seed is necessary for effective treatment planning. Well ionization chambers are used on site at therapy clinics to determine the air kerma strength of seeds. In this work, the response of the Standard Imaging HDR 1000 Plus well chamber to ambient pressure is examined using Monte Carlo calculations. The experimental work examining the response of this chamber as well as other chambers is presented in a companion paper. The Monte Carlo results show that for low‐energy photon sources, the application of the standard temperature pressure correction factor produces an over‐response at the reduced air densities/pressures corresponding to high elevations. With photon sources of 20 to 40 keV, the normalized corrected chamber response is as much as 10% to 20% over unity for air densities/pressures corresponding to an elevation of 3048 m (10000 ft) above sea level. At air densities corresponding to an elevation of 1524 m (5000 ft), the normalized ‐corrected chamber response is 5% to 10% over unity for these photon sources. With higher‐energy photon sources , the normalized corrected chamber response is near unity. For low‐energy sources of 0.25 to 0.50 MeV, the normalized ‐corrected chamber response is as much as 4% to 12% over unity for air densities/pressures corresponding to an elevation of 3048 m (10000 ft) above sea level. Higher‐energy sources have a normalized corrected chamber response near unity. Comparing calculated and measured chamber responses for common ‐ and ‐based brachytherapy seeds show agreement to within 2.7% and 1.9%, respectively. Comparing MCNP calculated chamber responses with EGSnrc calculated chamber responses show agreement to within 3.1% at photon energies of 20 to 40 keV. We conclude that Monte Carlo transport calculations accurately model the response of this well chamber. Further, applying the standard correction factor for this well chamber is insufficient in accounting for the change in chamber response with air pressure for low‐energy photon and low‐energy sources.Keywords
This publication has 21 references indexed in Scilit:
- Well-ionization chamber response relative to NIST air-kerma strength standard for prostate brachytherapy seedsMedical Physics, 2000
- Simulation of the response of the IG11 4πγ ionization chamber using GEANT Monte Carlo codeApplied Radiation and Isotopes, 2000
- AAPM's TG‐51 protocol for clinical reference dosimetry of high‐energy photon and electron beamsMedical Physics, 1999
- Code of practice for brachytherapy physics: Report of the AAPM Radiation Therapy Committee Task Group No. 56Medical Physics, 1997
- On the barometric formulaAmerican Journal of Physics, 1997
- ITS: the integrated TIGER series of electron/photon transport codes-Version 3.0IEEE Transactions on Nuclear Science, 1992
- A new re-entrant ionization chamber for the calibration of iridium-192 high dose rate sourcesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 1992
- Simulation of Dosimeter Response and Interface Effects1988
- Dose calibrator response to brachytherapy sources: A Monte Carlo and analytic evaluationMedical Physics, 1983
- Backscattering coefficient measurements of 15 to 60 keV electrons for solids at various angles of incidencePhysica Status Solidi (a), 1980