Absorbed Organ Doses and Effective Doses in Intraoral Radiography of Left Upper Molar of Adult Patient: A Computational Study
Open Access
Abstract
Objectives. To investigate effective doses and relevant absorbed organ doses for typical intraoral radiographic imaging procedures in adult patients. Variations of relevant absorbed organ doses and effective doses due to changes in imaging parameters and X-ray tube positioning were investigated. Conversion factors for estimating effective doses from incident air kerma (Ka,i) and kerma area product are proposed. Materials and Methods. Different imaging parameters and geometries obtained from previously published data from Serbia were simulated using Monte Carlo simulation and a computational anthropomorphic phantom. The relevant absorbed organ doses and effective doses for different exposure parameters and geometries were then compared to Ka,i and analyzed. Since an obsolete intraoral unit with a focus-to-skin distance of 10 cm, a 50 kVp beam and circular collimator with 5 cm diameter was found in the previous investigation, these imaging parameters were also included in the simulations. Results. For typical Ka,i for imaging protocols found in Serbia, effective doses for imaging of the left upper molar in an adult person were 2.9 µSv and 5.4 µSv for 60 kVp and 70 kVp beams, respectively. The highest absorbed organ doses were observed for the thyroid, eye lenses, salivary glands and brain, reaching up to 47.2 µGy, 71.4 µGy, 63.1 µGy and 26.2 µGy, respectively. For poorly optimized imaging protocols, these values were higher. Furthermore, for the obsolete intraoral units used, all doses were considerably higher and the typical effective dose was 16.8 µSv with thyroid dose of 105.3 µGy. Using rectangular instead of circular collimators decreased effective dose by 25% to 52.7% and absorbed organ doses by 43.3% to 84.5% in all simulated situations. Simulated positioning errors could change eye lens or thyroid absorbed organ dose by up to a factor of two when a rectangular collimator is used and up to four times when a circular collimator is used. Conclusions. Though the effective and absorbed organ doses from intraoral imaging are very low, this imaging can be further optimized by using optimal exposure parameters, rectangular collimators, and proper positioning of the X-ray tube.Keywords
Funding Information
- European Union (581-UNIOS-45)
This publication has 25 references indexed in Scilit:
- Lessons learnt from the recent EURADOS intercomparisons in computational dosimetryRadiation Measurements, 2022
- Monte Carlo calculation of organ and effective doses due to photon and neutron point sources and typical X-ray examinations: Results of an international intercomparison exerciseRadiation Measurements, 2022
- The use of dose quantities in radiological protection: ICRP publication 147 Ann ICRP 50(1) 2021Journal of Radiological Protection, 2021
- Quality assurance for the use of computational methods in dosimetry: activities of EURADOS Working Group 6 ‘Computational Dosimetry’Journal of Radiological Protection, 2021
- Diagnostic Imaging Techniques in Oral Diseases2019
- Dose area product in estimation of effective dose of the patients undergoing dental cone beam computed tomography examinationsJournal of Radiological Protection, 2018
- Modern dental imaging: a review of the current technology and clinical applications in dental practiceEuropean Radiology, 2010
- Title pageAnnals of the ICRP, 2007
- ENDF/B-VII.0: Next Generation Evaluated Nuclear Data Library for Nuclear Science and TechnologyNuclear Data Sheets, 2006
- Effective dose equivalent and effective dose: Comparison for common projections in oral and maxillofacial radiologyOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 2000