Closed-Loop Artificial Pancreas Using Subcutaneous Glucose Sensing and Insulin Delivery and a Model Predictive Control Algorithm: Preliminary Studies in Padova and Montpellier
Open Access
- 1 September 2009
- journal article
- research article
- Published by SAGE Publications in Journal of Diabetes Science and Technology
- Vol. 3 (5) , 1014-1021
Abstract
New effort has been made to develop closed-loop glucose control, using subcutaneous (SC) glucose sensing and continuous subcutaneous insulin infusion (CSII) from a pump, and a control algorithm. An approach based on a model predictive control (MPC) algorithm has been utilized during closed-loop control in type 1 diabetes patients. Here we describe the preliminary clinical experience with this approach. In Padova, two out of three subjects showed better performance with the closed-loop system compared to open loop. Altogether, mean overnight plasma glucose (PG) levels were 134 versus 111 mg/dl during open loop versus closed loop, respectively. The percentage of time spent at PG > 140 mg/dl was 45% versus 12%, while postbreakfast mean PG was 165 versus 156 mg/dl during open loop versus closed loop, respectively. Also, in Montpellier, two patients out of three showed a better glucose control during closed-loop trials. Avoidance of nocturnal hypoglycemic excursions was a clear benefit during algorithm-guided insulin delivery in all cases. This preliminary set of studies demonstrates that closed-loop control based entirely on SC glucose sensing and insulin delivery is feasible and can be applied to improve glucose control in patients with type 1 diabetes, although the algorithm needs to be further improved to achieve better glycemic control. Six type 1 diabetes patients (three in each of two clinical investigation centers in Padova and Montpellier), using CSII, aged 36 ± 8 and 48 ± 6 years, duration of diabetes 12 ± 8 and 29 ± 4 years, hemoglobin A1c 7.4% ± 0.1% and 7.3% ± 0.3%, body mass index 23.2 ± 0.3 and 28.4 ± 2.2 kg/m2, respectively, were studied on two occasions during 22 h overnight hospital admissions 2–4 weeks apart. A Freestyle Navigator® continuous glucose monitor and an OmniPod® insulin pump were applied in each trial. Admission 1 used open-loop control, while admission 2 employed closed-loop control using our MPC algorithm.Keywords
This publication has 23 references indexed in Scilit:
- Fully Automated Closed-Loop Insulin Delivery Versus Semiautomated Hybrid Control in Pediatric Patients With Type 1 Diabetes Using an Artificial PancreasDiabetes Care, 2008
- Detection of a Meal Using Continuous Glucose MonitoringDiabetes Care, 2008
- Prandial Insulin Dosing Using Run-to-Run ControlDiabetes Care, 2007
- Closed loop insulin delivery using implanted insulin pumps and sensors in type 1 diabetic patientsDiabetes Research and Clinical Practice, 2006
- Feasibility of Automating Insulin Delivery for the Treatment of Type 1 DiabetesDiabetes, 2006
- Model-based blood glucose control for type 1 diabetes via parametric programmingIEEE Transactions on Biomedical Engineering, 2006
- Run-to-run control of blood glucose concentrations for people with type 1 diabetes mellitusIEEE Transactions on Biomedical Engineering, 2006
- Interstitial fluid glucose dynamics during insulin-induced hypoglycaemiaDiabetologia, 2005
- Continuous glucose monitoring and closed‐loop systemsDiabetic Medicine, 2005
- Run-to-Run Control of Meal-Related Insulin DosingDiabetes Technology & Therapeutics, 2005