Hearing Aid Fitting by Prescription
Author:
- 1 January 1976
- journal article
- research article
- Published by Taylor & Francis in International Journal of Audiology
- Vol. 15 (2) , 163-173
Abstract
A procedure for the specification of hearing aid response characteristics is described. Thresholds of discomfort and most comfortable levels for narrow bands of noise centered at 0.25, 0.5, 1, 2, 3 and 4 kHz are determined. The thresholds of discomfort relate to the maximum power output of the hearing aid and the most comfortable levels determine the gain and frequency response that will be required. The data derived from these measurements are given to the hearing aid dispenser who provides a hearing aid that matches the prescribed response characteristics This technique requires less time than the conventional hearing aid evaluation procedure, and it provides the audiologist with measures of auditory function that can be directly related to hearing aid parametersKeywords
This publication has 3 references indexed in Scilit:
- Hearing Aid Gain Requirements in Sensori Neural Hearing LossBritish Journal of Audiology, 1973
- Speech Discrimination in Sensori-Neural Hearing Loss: Two Experiments on the Role of IntensityJournal of Speech and Hearing Research, 1966
- Tests for selection of hearing AIDSThe Laryngoscope, 1946