A multiple points method for 4D CT image sorting
Authors:
- Chiara GianoliCGChiara Gianoli
- Marco RiboldiMRMarco Riboldi
- Maria Francesca SpadeaMSMaria Francesca Spadea
- Laura Lavinia TravainiLTLaura Lavinia Travaini
- Mahila FerrariMFMahila Ferrari
- Riccardo MeiRMRiccardo Mei
- Roberto OrecchiaRORoberto Orecchia
- Guido BaroniGBGuido Baroni
- 10 January 2011
- journal article
- Published by Wiley in Medical Physics
- Vol. 38 (2) , 656-667
Abstract
Purpose: Artifacts affect 4D CT images due to breathing irregularities or incorrect breathing phase identification. The purpose of this study is the reduction of artifacts in sorted 4D CT images. The assumption is that the use of multiple respiratory related signals may reduce uncertainties and increase robustness in breathing phase identification.Methods: Multiple respiratory related signals were provided by infrared 3D localization of a configuration of markers placed on the thoracoabdominal surface. Multidimensional K‐means clustering was used for retrospective 4D CT image sorting, which was based on multiple marker variables, in order to identify clusters representing different breathing phases. The proposed technique was tested on computational simulations, phantom experimental acquisitions, and clinical data coming from two patients. Computational simulations provided a controlled and noise‐free condition for testing the clustering technique on regular and irregular breathing signals, including baseline drift, time variant amplitude, time variant frequency, and end‐expiration plateau. Specific attention was given to cluster initialization. Phantom experiments involved two moving phantoms fitted with multiple markers. Phantoms underwent 4D CT acquisition while performing controlled rigid motion patterns and featuring end‐expiration plateau. Breathing cycle period and plateau duration were controlled by means of weights leaned upon the phantom during repeated 4D CT scans. The implemented sorting technique was applied to clinical 4D CT scans acquired on two patients and results were compared to conventional sorting methods.Results: For computational simulations and phantom studies, the performance of the multidimensional clustering technique was evaluated by measuring the repeatability in identifying the breathing phase among adjacent couch positions and the uniformity in sampling the breathing cycle. When breathing irregularities were present, the clustering technique consistently improved breathing phase identification with respect to conventional sorting methods based on monodimensional signals. In patient studies, a qualitative comparison was performed between corresponding breathing phases of 4D CT images obtained by conventional sorting methods and by the described clustering technique. Artifact reduction was clearly observable on both data set especially in the lower lung region.Conclusions: The implemented multiple point method demonstrated the ability to reduce artifacts in 4D CT imaging. Further optimization and development are needed to make the most of the availability of multiple respiratory related variables and to extend the method to 4D CT‐PET hybrid scan.Keywords
This publication has 28 references indexed in Scilit:
- 4D treatment planning for scanned ion beamsRadiation Oncology, 2007
- The impact of temporal inaccuracies on 4DCT image qualityMedical Physics, 2007
- Design of 4D treatment planning target volumesInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2006
- Improving retrospective sorting of 4D computed tomography dataMedical Physics, 2006
- Four-dimensional image-based treatment planning: Target volume segmentation and dose calculation in the presence of respiratory motionInternational Journal of Radiation Oncology - Biology - Physics (IJROBP), 2005
- Fusion of respiration-correlated PET and CT scans: correlated lung tumour motion in anatomical and functional scansPhysics in Medicine & Biology, 2005
- Four‐dimensional radiotherapy planning for DMLC‐based respiratory motion trackingMedical Physics, 2005
- Four‐dimensional computed tomography: Image formation and clinical protocolMedical Physics, 2005
- Comparison of helical and cine acquisitions for 4D-CT imaging with multislice CTMedical Physics, 2005
- Acquiring a four-dimensional computed tomography dataset using an external respiratory signalPhysics in Medicine & Biology, 2002