Compatibility of endoclips in the gastrointestinal tract with magnetic resonance imaging
Open Access
- 6 October 2020
- journal article
- research article
- Published by Springer Nature in Scientific Reports
- Vol. 10 (1) , 1-7
Abstract
There are no clear guidelines on the compatibility between endoclips that remain in the gastrointestinal (GI) tract and magnetic resonance imaging (MRI). The purpose of this study was to investigate the effect of 3T (T) MRI on endoclips placed in excised pig tissues. Two types of endoclips were assessed: Olympus EZ (HX-610-135L) and QuickClip Pro (HZ-202LR). We assessed tissue damage or perforation and detachment of endoclips under 3T MRI magnetic field. We also evaluated the magnitude of force required to detach the endoclips from the porcine tissue. We measured the magnetic force acting on the Olympus EZ clips. QuickClip Pro clips were used as a control in this study. There was no tissue damage and no detachment of the endoclips (Olympus EZ and QuickClip Pro) during 3T MRI. The force required to detach the Olympus EZ clips ranged from 0.9 to 3.0 N. The translational magnetic force acting on the endoclips was 3.18 × 10–3 N. Ex vivo experiments showed that the magnetic field generated by 3T MRI did not cause tissue damage or perforation and did not detach the endoclips. Olympus EZ clips and QuickClip Pro clips in the GI tract appear to be safe during 3T MRI.Keywords
This publication has 10 references indexed in Scilit:
- Endoscopic Clip MRI Screening: A Canada-Wide Policy SurveyAmerican Journal of Roentgenology, 2017
- Are MR Compatible hemoclips safe after control of hemostasis?Endoscopy, 2013
- Radiofrequency heating and magnetic field interactions of fixed partial dentures during 3-tesla magnetic resonance imagingOral Surgery, Oral Medicine, Oral Pathology and Oral Radiology, 2013
- Evaluating long-term attachment of two different endoclips in the human gastrointestinal tractWorld Journal of Gastrointestinal Endoscopy, 2010
- Magnetic resonance imaging compatibility of endoclipsGastrointestinal Endoscopy, 2009
- Randomized controlled study of 3 different types of hemoclips for hemostasis of bleeding canine acute gastric ulcersGastrointestinal Endoscopy, 2006
- Orthodontic springs and auxiliary appliances: assessment of magnetic field interactions associated with 1.5 T and 3 T magnetic resonance systemsEuropean Radiology, 2006
- Magneto-Science2006
- Recording forces exerted on the bowel wall during colonoscopy: in vitro evaluationThe International Journal of Medical Robotics and Computer Assisted Surgery, 2005
- MR imaging of metallic implants and materials: a compilation of the literatureAmerican Journal of Roentgenology, 1988