A comparison between a white LED confocal imaging system and a conventional flash fundus camera using chromaticity analysis
Authors:
Preprint
- 15 January 2019
- preprint
- Published by Springer Nature in Research Square
Abstract
BACKGROUND: Conventional flash fundus camera captures color images that are over-saturated in the red channel, producing a retinal picture that looks washed-out and uniform. A white LED confocal device was recently introduced to provide a neutral-looking retinal image. With this study we aimed to evaluate the color rendering properties of the white LED confocal system and to compare it to a conventional flash fundus camera through the use of chromaticity analysis. METHODS: A fully automated white LED confocal device (Eidon, Centervue, Padova, Italy) was used to capture fundus images from healthy volunteers and patients with retinal diseases. The same pictures were acquired with a flash fundus camera (Triton, Topcon Corporation, Tokyo, Japan). All color images were evaluated with respect to the chromaticity. Color analysis was performed according to the image color signature. Color signature of an image was defined as the distribution of its pixels in the rgb chromaticity space. The descriptors used for the analysis are the average and variability of the barycenter positions, the average of the variability and the number of unique colors (NUC) of all signatures. RESULTS: Two hundred thirty-three color fundus photographs were acquired with each retinal camera. The images acquired by confocal white LED device demonstrated an average barycenter position (rgb = [0.448, 0.328, 0.224]) closer to the center of the chromaticity space, while conventional fundus camera provides images with a clear shift toward red at the expense of the blue and green (rgb = [0.574, 0.278, 0.148] (pCONCLUSIONS: Confocal white LED system provides well-balanced color images with a wider richness of color content compared to conventional flash fundus camera. The overall higher chromaticity of Eidon device may provide benefits in terms of discriminative power and diagnostic accuracy. KEYWORDS: Chromaticity; Confocal white LED system; conventional flash fundus camera; Eidon; Topcon.Keywords
All Related Versions
- Published version: BMC Ophthalmology, 19 (1), 1.