Pre-Clinical Evaluation of an Adult Extracorporeal Carbon Dioxide Removal System with Active Mixing for Pediatric Respiratory Support
- 13 December 2014
- journal article
- research article
- Published by SAGE Publications in The International Journal of Artificial Organs
- Vol. 37 (12) , 888-899
Abstract
The objective of this work was to conduct pre-clinical feasibility studies to determine if a highly efficient, active-mixing, adult extracorporeal carbon dioxide removal (ECCO2R) system can safely be translated to the pediatric population. The Hemolung Respiratory Assist System (RAS) was tested in vitro and in vivo to evaluate its performance for pediatric veno-venous applications. The Hemolung RAS operates at blood flows of 350-550 ml/min and utilizes an integrated pump-gas exchange cartridge with a membrane surface area of 0.59 m2 as the only component of the extracorporeal circuit. Both acute and seven-day chronic in vivo tests were conducted in healthy juvenile sheep using a veno-venous cannulation strategy adapted to the in vivo model. The Hemolung RAS was found to have gas exchange and pumping capabilities relevant to patients weighing 3-25 kg. Seven-day animal studies in juvenile sheep demonstrated that veno-venous extracorporeal support could be used safely and effectively with no significant adverse reactions related to device operation.Keywords
This publication has 15 references indexed in Scilit:
- A Novel Extracorporeal CO2 Removal SystemChest, 2013
- Respiratory dialysis with an active-mixing extracorporeal carbon dioxide removal system in a chronic sheep studyIntensive Care Medicine, 2012
- In Vitro Performance Testing of a Pediatric Oxygenator With an Integrated Pulsatile PumpASAIO Journal, 2012
- In Vitro Pumping Performance Evaluation of the Ension Pediatric Cardiopulmonary Assist System for Venoarterial and Venovenous ECMOCardiovascular Engineering and Technology, 2012
- NeonatOx: A Pumpless Extracorporeal Lung Support for Premature NeonatesArtificial Organs, 2011
- A Miniaturized Extracorporeal Membrane Oxygenator with Integrated Rotary Blood Pump: Preclinical In Vivo TestingASAIO Journal, 2011
- The Aachen MiniHLM—A Miniaturized Heart‐Lung Machine for Neonates With an Integrated Rotary Blood PumpArtificial Organs, 2010
- Functional and Biocompatibility Performances of an Integrated Maglev Pump‐OxygenatorArtificial Organs, 2009
- Red Blood Cell Damage by Shear StressBiophysical Journal, 1972
- A simple empirical formula for calculating approximate surface area in children.Archives of Disease in Childhood, 1966