Digital Cell Counting Device Integrated with a Single-Cell Array
Open Access
- 13 February 2014
- journal article
- research article
- Published by Public Library of Science (PLoS) in PLOS ONE
- Vol. 9 (2) , e89011
Abstract
In this paper, we present a novel cell counting method accomplished using a single-cell array fabricated on an image sensor, complementary metal oxide semiconductor sensor. The single-cell array was constructed using a microcavity array, which can trap up to 7,500 single cells on microcavities periodically arranged on a plane metallic substrate via the application of a negative pressure. The proposed method for cell counting is based on shadow imaging, which uses a light diffraction pattern generated by the microcavity array and trapped cells. Under illumination, the cell-occupied microcavities are visualized as shadow patterns in an image recorded by the complementary metal oxide semiconductor sensor due to light attenuation. The cell count is determined by enumerating the uniform shadow patterns created from one-on-one relationships with single cells trapped on the microcavities in digital format. In the experiment, all cell counting processes including entrapment of non-labeled HeLa cells from suspensions on the array and image acquisition of a wide-field-of-view of 30 mm2 in 1/60 seconds were implemented in a single integrated device. As a result, the results from the digital cell counting had a linear relationship with those obtained from microscopic observation (r2 = 0.99). This platform could be used at extremely low cell concentrations, i.e., 25–15,000 cells/mL. Our proposed system provides a simple and rapid miniaturized cell counting device for routine laboratory use.Keywords
This publication has 21 references indexed in Scilit:
- Monitoring of benzene-induced hematotoxicity in mice by serial leukocyte counting using a microcavity arrayBiosensors and Bioelectronics, 2013
- Rapid quantification of pathogenic fungi by Cellometer image-based cytometryJournal of Microbiological Methods, 2012
- Leukocyte counting from a small amount of whole blood using a size‐controlled microcavity arrayBiotechnology and Bioengineering, 2012
- Overview of single-cell analyses: microdevices and applicationsLab on a Chip, 2010
- Note: Microelectromechanical systems Coulter counter for cell monitoring and countingReview of Scientific Instruments, 2010
- Size-Selective Microcavity Array for Rapid and Efficient Detection of Circulating Tumor CellsAnalytical Chemistry, 2010
- Validation of cell density and viability assays using Cedex automated cell counterBiologicals, 2010
- High-Density Microcavity Array for Cell Detection: Single-Cell Analysis of Hematopoietic Stem Cells in Peripheral Blood Mononuclear CellsAnalytical Chemistry, 2009
- Method for the Accurate Preparation of Cell-Spiking StandardsAnalytical Chemistry, 2008
- Performance evaluation of the Z2 coulter counter for WBC and RBC countingInternational Journal of Laboratory Hematology, 2007