Digital morphology compared to the optical microscope: A validation study on reporting bone marrow aspirates
- 8 February 2024
- journal article
- research article
- Published by Wiley in International Journal of Laboratory Hematology
- Vol. 46 (3) , 474-480
Abstract
Introduction: This study aims to evaluate the effectiveness and reliability of the utilization for clinical reporting of the evaluation of digital images of bone marrow aspirates by morphologists and their comparability with the classic microscopic morphological evaluation.Methods: We scanned 180 consecutive bone marrow needle aspirates smears using the “Metafer4 VSlide” whole slide imaging (WSI) digital scanning system. We evaluated the statistical comparability and the risk of bias of the microscopic readings with those performed on the screen on the digitized medullary images.Results: The evaluation of cellularity on the screen was equivalent, with a higher frequency of “normal” than the analysis of digital preparations. The means and medians of the percentage values obtained on the different cell populations with the microscopic and digital reading were comparable as the main categories are concerned, with an average difference equal to 0 for the neutrophilic and eosinophilic granulocytic series, at −0.2% for the total myeloid cells, at 1.2% for the erythroid series, at −0.4% for the lymphocytes and at −0.4% for the blasts. Dysplastic features were consistently identified in 69/71 cell lineages.Conclusion: Our study demonstrated that screen evaluation of digitized bone marrow needle aspirates provides quantitative and qualitative results comparable to traditional microscopic analysis of the corresponding slide smears. Digital images offer significant benefits in reducing the workload of experienced operators, reproducibility and sharing of observations, and image preservation. Even in routine diagnostic activities, their use does not alter the quality of the results obtained in evaluating bone marrow needle aspirates.Keywords
This publication has 26 references indexed in Scilit:
- Comment on: 'Should clinical hematologists put their microscopes on eBay?'Haematologica, 2014
- Interobserver variance in myelodysplastic syndromes with less than 5 % bone marrow blasts: unilineage vs. multilineage dysplasia and reproducibility of the threshold of 2 % blastsAnnals of Hematology, 2014
- Should clinical hematologists put their microscopes on eBay?Haematologica, 2014
- Morphological classification of the myelodysplastic syndromes: how much more education of diagnosticians is necessary?Haematologica, 2013
- Implications of discrepancy in morphologic diagnosis of myelodysplastic syndrome between referral and tertiary care centersBlood, 2011
- Confidence in Altman–Bland plots: A critical review of the method of differencesClinical and Experimental Pharmacology and Physiology, 2010
- ICSH guidelines for the standardization of bone marrow specimens and reportsInternational Journal of Laboratory Hematology, 2008
- Performance Evaluation of the CellaVision DM96 SystemAmerican Journal of Clinical Pathology, 2005
- The use of digital ‘virtual slides’ in the quality assessment of haematological morphology: results of a pilot exercise involving UK NEQAS(H) participantsBritish Journal of Haematology, 2005
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986