Comparison of Near-Wall Hemodynamic Parameters in Stented Artery Models
- 27 April 2009
- journal article
- research article
- Published by ASME International in Journal of Biomechanical Engineering
- Vol. 131 (6) , 061006
Abstract
Four commercially available stent designs (two balloon expandable—Bx Velocity and NIR, and two self-expanding—Wallstent and Aurora) were modeled to compare the near-wall flow characteristics of stented arteries using computational fluid dynamics simulations under pulsatile flow conditions. A flat rectangular stented vessel model was constructed and simulations were carried out using rigid walls and sinusoidal velocity input (nominal wall shear stress of ). Mesh independence was determined from convergence of the axial wall shear stress (WSS) along the length of the stented model. The flow disturbance was characterized and quantified by the distributions of axial and transverse WSS, WSS gradients, and flow separation parameters. Normalized time-averaged effective WSS during the flow cycle was the smallest for the Wallstent compared with the others ( for the Bx Velocity stent, for the Aurora stent, and for the NIR stent). Regions of low mean WSS and elevated WSS gradients were also the largest for the Wallstent compared with the others. WSS gradients were the largest near the struts and remained distinctly nonzero for most of the region between the struts for all stent designs. Fully recirculating regions (as determined by separation parameter) were the largest for the Bx Velocity stent compared with the others. The most hemodynamically favorable stents from our computational analysis were the Bx Velocity and NIR stents, which were slotted-tube balloon-expandable designs. Since clinical data indicate lower restenosis rates for the Bx Velocity and NIR stents compared with the Wallstent, our data suggest that near-wall hemodynamics may predict some aspects of in vivo performance. Further consideration of biomechanics, including solid mechanics, in stent design is warranted.
Keywords
This publication has 31 references indexed in Scilit:
- Effects of different stent designs on local hemodynamics in stented arteriesJournal of Biomechanics, 2008
- Drug-eluting stents and stent thrombosis: a cause for concern?Coronary Artery Disease, 2006
- Spatial Distribution of Platelet Deposition in Stented Arterial Models Under Physiologic FlowAnnals of Biomedical Engineering, 2005
- Circumferential vascular deformation after stent implantation alters wall shear stress evaluated with time-dependent 3D computational fluid dynamics modelsJournal of Applied Physiology, 2005
- Hemodynamic Parameters and Early Intimal Thickening in Branching Blood VesselsCritical Reviews in Biomedical Engineering, 2001
- Computational haemodynamics analysis and comparison study of arterio-venous graftsJournal of Medical Engineering & Technology, 2000
- Pathobiologic Responses to Stenting 11Supported in part by grants from the National Institutes of Health (GM/HL49039 and HL03104), the Burroughs Wellcome Fund for Experimental Therapeutics, Durham, North Carolina, and the Whitaker Foundation, Rosslyn, Virginia.The American Journal of Cardiology, 1998
- Fluid wall shear stress measurements in a model of the human abdominal aorta: oscillatory behavior and relationship to atherosclerosisAtherosclerosis, 1994
- Spatial and temporal variations of wall shear stress within an end-to-side arterial anastomosis modelJournal of Biomechanics, 1993
- Graft geometry and venous intimal-medial hyperplasia in arteriovenous loop graftsJournal of Vascular Surgery, 1990