A simplified method of preparing phosphoric acid for stable isotope analyses of carbonates

Authors:

Abstract
In order to produce CO 2 for stable isotope analyses ( δ 18 O and δ 13 C), carbonate samples are commonly digested in phosphoric acid. The acid recipe here presented is based on phase shifting crystalline orthophosphoric acid of pro‐analysis quality to a liquid state through heating, followed by pre‐vacuum treatment during a start‐up procedure before mass analyses for common acid bath preparation, or adding a small amount of phosphoric pentoxide for single drop equipments, respectively. This methodology results in a final acid concentration of 104%. Copyright © 2005 John Wiley & Sons, Ltd.

This publication has 6 references indexed in Scilit: