Abstract
Aims: Paravalvular regurgitation (PVR) is frequently observed following Transcatheter Aortic Valve Replacement (TAVR). Periprocedural monitoring of invasive hemodynamics has shown promise for diagnosis of PVR, but automated software options are lacking. We aimed to develop a rule-based algorithm for automated assessment of hemodynamic indices of PVR, and evaluate its construct validity and discriminatory value for cardiac magnetic resonance (CMR)-derived relevant PVR compared to standard manual hemodynamic assessment. Methods and results: Left ventricular and aortic pressures were invasively measured during TAVR using fluid-filled pigtail catheters. To evaluate construct validity of automated vs. manual assessment of invasive hemodynamics, we compared (i) proportion of cardiac cycles affected by arrhythmias/noise, (ii) pressure gradients, and (iii) PVR indices. Additionally, we compared the discriminatory value of automatically and manually determined PVR indices for CMR-determined relevant PVR at 30-days. In total, 77 patients were enrolled (664 cardiac cycles). Automated filtering of cardiac cycles affected by arrhythmias/noise had a high sensitivity (95.2%) and specificity (86.4%). In addition, excellent agreement was observed between automated and manual computation of mean gradients pre- and post-TAVR [39.3 ± 12.1 vs. 37.5 ± 11.9 mmHg, intra-class correlation coefficient (ICC): 0.916; 1.92 ± 5.87 vs. 1.14 ± 5.89, ICC: 0.957, respectively], and PVR indices [diastolic delta (DD): 41.7 ± 12.4 vs. 40.6 ± 12.3 mmHg, ICC: 0.982, respectively]. Automated and manual assessment of DD showed comparable discriminatory value for relevant PVR [area under the curve (AUC): 0.81 vs. 0.80, respectively]. Conclusion: Rule-based, automated assessment of hemodynamic indices of PVR showed excellent construct validity and discriminatory value for CMR-determined relevant PVR, supporting its use for real-time evaluation and risk stratification in TAVR patients.Keywords
Funding Information
- APPOSE
- Abbott Laboratories
This publication has 29 references indexed in Scilit:
- Five-Year Outcomes of Transcatheter or Surgical Aortic-Valve ReplacementNew England Journal of Medicine, 2020
- Incidence, Management, and Associated Clinical Outcomes of New-Onset Atrial Fibrillation Following Transcatheter Aortic Valve ReplacementJACC: Cardiovascular Interventions, 2018
- Cardiovascular Magnetic Resonance to Evaluate Aortic Regurgitation After Transcatheter Aortic Valve ReplacementJACC, 2016
- Hemodynamic Assessment of Aortic Regurgitation After Transcatheter Aortic Valve ReplacementJACC: Cardiovascular Interventions, 2016
- Pre-Procedural Hemodynamic Status Improves the Discriminatory Value of the Aortic Regurgitation Index in Patients Undergoing Transcatheter Aortic Valve ReplacementJACC: Cardiovascular Interventions, 2016
- 5-year outcomes of transcatheter aortic valve replacement compared with standard treatment for patients with inoperable aortic stenosis (PARTNER 1): a randomised controlled trialThe Lancet, 2015
- 5-year outcomes of transcatheter aortic valve replacement or surgical aortic valve replacement for high surgical risk patients with aortic stenosis (PARTNER 1): a randomised controlled trialThe Lancet, 2015
- Novel hemodynamic index for assessment of aortic regurgitation after transcatheter aortic valve replacementCatheterization and Cardiovascular Interventions, 2015
- From pressure overload to volume overload: Aortic regurgitation after transcatheter aortic valve implantationAmerican Heart Journal, 2012
- Aortic Regurgitation Index Defines Severity of Peri-Prosthetic Regurgitation and Predicts Outcome in Patients After Transcatheter Aortic Valve ImplantationJACC, 2012