Detectable sizes of shallow flaws on thin copper tubes were experimentally investigated using an eddy-current-based nondestructive inspection system with a high-Tc superconducting quantum interference device (SQUID) gradiometer. At an excitation field of 5.6 µT at 3 kHz, a magnetic anomaly due to the shallowest flaw of 10 µm depth, 100 µm width, and 15 mm length on the surface of a tube specimen of 6.35 mm outer diameter and 0.8 mm thickness was successfully measured. Signal amplitude due to a flaw was proportional to effective flaw volume that made an induced eddy current detour. The experimental results indicate that the detectable volume of a surface flaw on a copper tube should be 1×106 µm3 with some increase in signal-to-noise ratio.