Recording evoked potentials during deep brain stimulation: development and validation of instrumentation to suppress the stimulus artefact
Authors:
- 18 April 2012
- journal article
- research article
- Published by IOP Publishing in Journal of Neural Engineering
- Vol. 9 (3) , 036004
Abstract
The clinical efficacy of deep brain stimulation (DBS) for the treatment of movement disorders depends on the identification of appropriate stimulation parameters. Since the mechanisms of action of DBS remain unclear, programming sessions can be time consuming, costly and result in sub-optimal outcomes. Measurement of electrically evoked compound action potentials (ECAPs) during DBS, generated by activated neurons in the vicinity of the stimulating electrode, could offer insight into the type and spatial extent of neural element activation and provide a potential feedback signal for the rational selection of stimulation parameters and closed-loop DBS. However, recording ECAPs presents a significant technical challenge due to the large stimulus artefact, which can saturate recording amplifiers and distort short latency ECAP signals. We developed DBS-ECAP recording instrumentation combining commercial amplifiers and circuit elements in a serial configuration to reduce the stimulus artefact and enable high fidelity recording. We used an electrical circuit equivalent model of the instrumentation to understand better the sources of the stimulus artefact and the mechanisms of artefact reduction by the circuit elements. In vitro testing validated the capability of the instrumentation to suppress the stimulus artefact and increase gain by a factor of 1000 to 5000 compared to a conventional biopotential amplifier. The distortion of mock ECAP (mECAP) signals was measured across stimulation parameters, and the instrumentation enabled high fidelity recording of mECAPs with latencies of only 0.5 ms for DBS pulse widths of 50 to 100 µs/phase. Subsequently, the instrumentation was used to record in vivo ECAPs, without contamination by the stimulus artefact, during thalamic DBS in an anesthetized cat. The characteristics of the physiological ECAP were dependent on stimulation parameters. The novel instrumentation enables high fidelity ECAP recording and advances the potential use of the ECAP as a feedback signal for the tuning of DBS parameters.Keywords
This publication has 56 references indexed in Scilit:
- Current-controlled deep brain stimulation reduces in vivo voltage fluctuations observed during voltage-controlled stimulationClinical Neurophysiology, 2010
- Deep Brain Stimulation Alleviates Parkinsonian Bradykinesia by Regularizing Pallidal ActivityJournal of Neurophysiology, 2010
- Impedance characteristics of deep brain stimulation electrodesin vitroandin vivoJournal of Neural Engineering, 2009
- Experimental and theoretical characterization of the voltage distribution generated by deep brain stimulationExperimental Neurology, 2009
- The influence of reactivity of the electrode–brain interface on the crossing electric current in therapeutic deep brain stimulationNeuroscience, 2008
- Tremor varies as a function of the temporal regularity of deep brain stimulationNeuroReport, 2008
- Long-term evaluation of deep brain stimulation of the thalamusJournal of Neurosurgery, 2006
- Tissue and electrode capacitance reduce neural activation volumes during deep brain stimulationClinical Neurophysiology, 2005
- Dielectric properties of porcine brain tissue in the transition from life to death at frequencies from 800 to 1900 MHzBioelectromagnetics, 2003
- Computation of dielectric constants of solvent mixtures and electrolyte solutionsFluid Phase Equilibria, 2001