Radiofrequency heating of metallic dental devices during 3.0 T MRI
Authors:
- M HasegawaMHM Hasegawa
- K MiyataKMK Miyata
- Y AbeYAY Abe
- T IshigamiTIT Ishigami
- 1 May 2013
- journal article
- Published by Oxford University Press (OUP) in Dentomaxillofacial Radiology
- Vol. 42 (5) , 20120234
Abstract
Objectives:To estimate the risk of injury from radiofrequency (RF) heating of metallic dental devices in use during 3.0 T MRI.Methods:The whole-body specific absorption rate (WB-SAR) was calculated on the basis of saline temperature elevation under the maximum RF irradiation for 15 min to determine the operation parameters for the heating test. The temperature changes of three types of three-unit bridges, a full-arch fixed dental prosthesis and an orthodontic appliance in use during MRI with a 3.0 T MR system (Magnetom® Verio; Siemens AG, Erlangen, Germany) were then tested in accordance with the American Society for Testing and Materials F2182-09 standardized procedure under the maximum RF heating during 15 min RF irradiation.Results:The system console-predicted WB-SAR was approximately 1.4 W kg−1 and that measured with a saline phantom was 2.1 W kg−1. In the assessment of RF heating, the highest temperature increase was +1.80 °C in the bridges, +1.59 °C in the full-arch fixed dental prosthesis and +2.61...Keywords
This publication has 25 references indexed in Scilit:
- Spatial distribution of RF‐induced E‐fields and implant heating in MRIMagnetic Resonance in Medicine, 2008
- Clinical advantages of 3.0T MRI over 1.5TEuropean Journal of Radiology, 2008
- Variability in RF‐induced heating of a deep brain stimulation implant across MR systemsJournal of Magnetic Resonance Imaging, 2006
- The effect of box preparation on the strength of glass fiber–reinforced composite inlay-retained fixed partial denturesThe Journal of Prosthetic Dentistry, 2005
- Neurostimulation systems: Assessment of magnetic field interactions associated with 1.5‐ and 3‐Tesla MR systemsJournal of Magnetic Resonance Imaging, 2004
- Evaluation of specific absorption rate as a dosimeter of MRI‐related implant heatingJournal of Magnetic Resonance Imaging, 2004
- In vivo findings associated with heat generation during thermomechanical compactionof gutta‐percha. Part II. Histological response to temperature elevation on the external surface of the rootInternational Endodontic Journal, 1990
- A potential MRI hazard: forces on dental magnet keepersJournal of Oral Rehabilitation, 1990
- Temperature threshold levels for heat-induced bone tissue injury: A vital-microscopic study in the rabbitThe Journal of Prosthetic Dentistry, 1983
- Microscopic Observations on Living Dental Pulp II. The Effect of Thermal Irritants on the Circulation of the Pulp in the Lower Rat IncisorActa Odontologica Scandinavica, 1958