Analyzing the characteristics of 6 MV photon beam at low monitor unit settings
Open Access
- 1 January 2016
- journal article
- research article
- Published by Medknow in Journal of Medical Physics
- Vol. 41 (1) , 34-7
Abstract
Analyzing the characteristics of a low monitor unit (MU) setting is essential, particularly for intensity-modulated techniques. Intensity modulation can be achieved through intensity-modulated radiotherapy (IMRT) or volumetric-modulated arc therapy (VMAT). There is possibility for low MUs in the segments of IMRT and VMAT plans. The minimum MU/segment must be set by the physicist in the treatment planning system at the time of commissioning. In this study, the characteristics such as dose linearity, stability, flatness, and symmetry of 6 MV photon beam of a Synergy linear accelerator at low MU settings were investigated for different dose rates. The measurements were performed for Synergy linear accelerator using a slab phantom with a FC65-G chamber and Profiler 2. The MU linearity was studied for 1–100 MU using a field size of 10 cm ×10 cm. The linearity error for 1 MU was 4.2%. Flatness of the beam was deteriorated in 1 MU condition. The beam stability and symmetry was well within the specification. Using this study, we conclude that the treatment delivered with <3 MU may result in uncertainty in dose delivery. To ensure the correct dose delivery with less uncertainty, it is recommended to use ≥3 MU as the minimum MU per segment in IMRT and VMAT plans.Keywords
This publication has 11 references indexed in Scilit:
- Characteristics of flattening filter free beams at low monitor unit settingsMedical Physics, 2013
- A dosimetric comparative study: Volumetric modulated arc therapy vs intensity-modulated radiation therapy in the treatment of nasal cavity carcinomasMedical Dosimetry, 2013
- Clinical experience transitioning from IMRT to VMAT for head and neck cancerMedical Dosimetry, 2013
- Multi-institutional comparison of volumetric modulated arc therapy vs. intensity-modulated radiation therapy for head-and-neck cancer: a planning studyRadiation Oncology, 2013
- Beam characteristics of megavoltage beams at low monitor unit settingsPhysica Medica, 2011
- Dosimetric characteristics of linear accelerator photon beams with small monitor unit settingsMedical Physics, 2008
- Linearity of the Dose Monitor System at Low Monitor UnitsStrahlentherapie und Onkologie, 2007
- Dose linearity and uniformity of Siemens ONCOR impression plus linear accelerator designed for step-and-shoot intensity-modulated radiation therapyJournal of Medical Physics, 2007
- Performance characterization of siemens primus linear accelerator under small monitor unit and small segments for the implementation of step-and-shoot intensitymodulated radiotherapyJournal of Medical Physics, 2006
- Dose Delivery Accuracy of Therapeutic Photon and Electron Beams at Low Monitor Unit SettingsStrahlentherapie und Onkologie, 2005