Spectral Mismatch Effect of Ultraviolet Radiometers in Actual UV‐C Measurement
- 4 January 2023
- journal article
- research article
- Published by Wiley in Photochemistry and Photobiology
- Vol. 99 (5) , 1240-1247
Abstract
The management of radiant exposure to ultraviolet (UV) radiation, especially in the wavelength range from 100 nm to 280 nm (i.e. UV-C), is important for virus inactivation or photobiological safety. Recently, many commercial UV radiometers have been used to measure UV-C irradiance for industrial and public applications. The accuracy of the four types of commercial UV radiometers was investigated by comparing the reference irradiance values obtained from the spectral irradiance standard. It was found that the displayed values of the UV radiometers have discrepancies, such that the measured value can be more than twice the actual value in a certain case. The spectral mismatch between the calibration and test sources is a major factor in the discrepancies in the UV-C measurements. With spectral mismatch correction, most corrected values show a tendency to improve the result to approaching the reference values within 20%. Users need to provide spectral information about the source and radiometer used for UV-C measurement.Keywords
This publication has 18 references indexed in Scilit:
- Chronic irradiation with 222-nm UVC light induces neither DNA damage nor epidermal lesions in mouse skin, even at high dosesPLOS ONE, 2018
- Disinfection and healing effects of 222-nm UVC light on methicillin-resistant Staphylococcus aureus infection in mouse woundsJournal of Photochemistry and Photobiology B: Biology, 2018
- Germicidal Efficacy and Mammalian Skin Safety of 222-nm UV LightRadiation Research, 2017
- The CCPR K2.c key comparison of spectral responsivity from 200 nm to 400 nmMetrologia, 2014
- 207-nm UV Light - A Promising Tool for Safe Low-Cost Reduction of Surgical Site Infections. I: In Vitro StudiesPLOS ONE, 2013
- ICNIRP STATEMENT—PROTECTION OF WORKERS AGAINST ULTRAVIOLET RADIATIONHealth Physics, 2010
- Final report on CCPR K1-a: Spectral irradiance from 250 nm to 2500 nmMetrologia, 2006
- Calibration of broadband UV radiometers methodology and uncertainty evaluationMetrologia, 2003
- Sources of error in UV radiation measurementsJournal of Research of the National Institute of Standards and Technology, 2001
- Characterization and calibration of broadband ultraviolet radiometersMetrologia, 2000