Comparison of the performance of the IDEXX SediVue Dx® with manual microscopy for the detection of cells and 2 crystal types in canine and feline urine
Open Access
- 3 December 2018
- journal article
- research article
- Published by Oxford University Press (OUP) in Journal of Veterinary Internal Medicine
- Vol. 33 (1) , 167-177
Abstract
Background: Microscopic evaluation of urine is inconsistently performed in veterinary clinics. The IDEXX SediVue Dx® Urine Sediment Analyzer (SediVue) recently was introduced for automated analysis of canine and feline urine and may facilitate performance of urinalyses in practice. Objective: Compare the performance of the SediVue with manual microscopy for detecting clinically relevant numbers of cells and 2 crystal types. Samples: Five-hundred thirty urine samples (82% canine, 18% feline). Methods: For SediVue analysis (software versions [SW] 1.0.0.0 and 1.0.1.3), uncentrifuged urine was pipetted into a cartridge. Images were captured and processed using a convolutional neural network algorithm. For manual microscopy, urine was centrifuged to obtain sediment. To determine sensitivity and specificity of the SediVue compared with manual microscopy, thresholds were set at ≥5/high power field (hpf) for red blood cells (RBC) and white blood cells (WBC) and ≥1/hpf for squamous epithelial cells (sqEPI), non-squamous epithelial cells (nsEPI), struvite crystals (STR), and calcium oxalate dihydrate crystals (CaOx Di). Results: The sensitivity of the SediVue (SW1.0.1.3) was 85%-90% for the detection of RBC, WBC, and STR; 75% for CaOx Di; 71% for nsEPI; and 33% for sqEPI. Specificity was 99% for sqEPI and CaOx Di; 87%-90% for RBC, WBC, and nsEPI; and 84% for STR. Compared to SW1.0.0.0, SW1.0.1.3 had increased sensitivity but decreased specificity. Performance was similar for canine versus feline and fresh versus stored urine samples. Conclusions and Clinical Importance: The SediVue exhibits good agreement with manual microscopy for the detection of most formed elements evaluated, but improvement is needed for epithelial cells.Keywords
Funding Information
- IDEXX Laboratories, Inc
- IDEXX Laboratories, Inc
This publication has 18 references indexed in Scilit:
- Comparison of Fully Automated Urine Sediment Analyzers H800‐FUS100 and Labumat‐Urised with Manual MicroscopyJournal of Clinical Laboratory Analysis, 2013
- Urine sediment analysis: Analytical and diagnostic performance of sediMAX® — A new automated microscopy image-based urine sediment analyserClinica Chimica Acta, 2010
- Comparison of LabUMat‐with‐UriSed and iQ®200 fully automatic urine sediment analysers with manual urine analysisBiotechnology and Applied Biochemistry, 2009
- Interobserver Reliability of Urine Sediment InterpretationClinical Journal of the American Society of Nephrology, 2009
- Clinical laboratory automated urinalysis: comparison among automated microscopy, flow cytometry, two test strips analyzers, and manual microscopic examination of the urine sedimentsJournal of Clinical Laboratory Analysis, 2008
- Automated urinalysis: first experiences and a comparison between the Iris iQ200 urine microscopy system, the Sysmex UF-100 flow cytometer and manual microscopic particle countingClinical Chemistry and Laboratory Medicine (CCLM), 2007
- Analytical performance of the Iris iQ200 automated urine microscopy analyzerClinica Chimica Acta, 2006
- Safely Reducing Manual Urine Microscopy Analyses by Combining Urine Flow Cytometer and Strip ResultsAmerican Journal of Clinical Pathology, 2001
- Automation of Urine Sediment Examination: a Comparison of the Sysmex UF-100 Automated Flow Cytometer with Routine Manual Diagnosis (Microscopy, Test Strips, and Bacterial Culture)Clinical Chemistry and Laboratory Medicine (CCLM), 1999
- The Automation of Sediment Urinalysis Using a New Urine Flow Cytometer (UF-100™)Clinical Chemistry and Laboratory Medicine (CCLM), 1998