Abstract
Introduction Neonatal hypothermia in low-resource settings is prevalent and closely associated with high morbidity and mortality. We examined if an easy-to-read temperature detector device improves health outcomes. Methods In a descriptive study, 1009 admissions to a neonatal ward in a tertiary care hospital in Lilongwe, Malawi, were analysed and divided into a baseline and a trial group. The data of 531 newborns with standard care (SC) before the trial were compared with 478 newborns during the implementation of the device (device care=DC). Staff and caregivers were trained on using the device and how to react in case of hypothermia. Data were collected from patient files, device documentation sheets, interviews and focus group discussions. Hypothermia was defined as a body temperature Results During the trial, body temperatures throughout the hospital stay were significantly more often obtained (pConclusion With more frequent temperature checks, infants with lower birth weight possibly benefited from implementing an easy-to-read continuous temperature indicator, but hypothermia rates remained high. Our data and experiences reveal structural, communicational and consistency/interpretation deficits. Although specifically designed for low-resource settings, the implementation of the device needs a well-working and structured environment, especially regarding staff and caregiver communication.
Funding Information
  • Institut für Medizinische Informationsverarbeitung Biometrie und Epidemiologie
  • The Else-Kröner-Fresenius Stiftung, Germany

This publication has 29 references indexed in Scilit: