Influence of the Compliance of the Pump Housing and Cannulas of a Paracorporeal Pneumatic Ventricular Assist Device on Transient Pressure Characteristics
Authors:
- 1 June 1995
- journal article
- Published by Wiley in Artificial Organs
- Vol. 19 (6) , 525-534
Abstract
The dependence of transient pressure characteristics of a ventricular assist device (VAD) on the compliance of its housing and cannulas was investigated in a mock circulation. The peak rate of change of pressure (dP/dtmax) values was greater in the cannulas than other compartments and was associated with valve closure‐induced pressure oscillations. When cannula compliance was increased from 0.0057 to 0.0129 cm3/mm Hg, these values decreased by ˜20%, and outflow cannula pressure oscillation frequency decreased from 17.5 Hz by 35%. This trend was also apparent in the inflow. A VAD housing compliance increase from 0.0162 to 0.0483 cm3/mm Hg caused a dP/dtmax decrease of 30% in both the blood chamber and the outflow cannula. The effect of this change on the inflow was weaker implying that housing absorbs the energy associated with outflow deceleration more effectively than the inflow. These findings suggest that increasing VAD housing and cannulas compliance can improve hydrodynamic performance.Keywords
This publication has 8 references indexed in Scilit:
- Pressure development within a sac-type pneumatically driven ventricular assist deviceJournal of Biomechanics, 1994
- A synthetic leaflet heart valve with improved opening characteristicsMedical Engineering & Physics, 1994
- Experimental investigation of the motions of the pumping diaphragm within a sac-type pneumatically driven ventricular assist deviceJournal of Biomechanics, 1994
- Experimental investigation of unsteady flow behaviour within a sac-type ventricular assist device (VAD)Journal of Biomechanics, 1993
- Development of mock circulation models for the assessment of counterpulsation systemsCardiovascular Research, 1991
- The Fluid Mechanics of a Sac-Type Ventricular Assist DeviceThe International Journal of Artificial Organs, 1990
- Comparative Evaluation of Disk- and Tri Leaflet Valves in Left-Ventricular Assist Devices (LVAD)The International Journal of Artificial Organs, 1988
- An artificial arterial system for pumping hearts.Journal of Applied Physiology, 1971