Effects of Dual-Energy Technique on Radiation Exposure and Image Quality in Pediatric Body CT

Authors:

Abstract
OBJECTIVE. The purpose of this study was to assess the effects of dual-energy CT (DECT) on radiation exposure and image quality in pediatric body CT. MATERIALS AND METHODS. This retrospective study included 79 children (median age, 10.1 years; range, 12 days–18 years) who underwent thoracic or abdominal-pelvic CT or CT angiography with dual-energy technique between October 2014 and March 2015. The delivered volume CT dose index (CTDIvol) from DECT was recorded and compared with the estimated CTDIvol had the patient undergone scanning with a standard single-energy CT (SECT) protocol. Size-specific dose estimates were calculated for both DECT and SECT. Image quality was subjectively scored (scale, 1–4). For 16 of 79 patients who underwent both DECT and SECT, image contrast and noise were measured and contrast-to-noise ratio calculated. Parametric and nonparametric testing of independent and paired samples was performed. RESULTS. For all 79 studies, actual median CTDIvol and size-specific dose estimate were ...