Clinical and Electrophysiological Properties of a New Temporary Pacemaker Lead after Open‐Heart Surgery
Authors:
- 1 July 1982
- journal article
- Published by Wiley in Pacing and Clinical Electrophysiology
- Vol. 5 (4) , 600-606
Abstract
A new temporary pacemaker lead, Medtronic 6400, with a solid defined electrode surface area of 7.5 mm2, has been impianted in 50 patients after open‐heart surgery. One electrode was inserted intramyocardially on the right ventricle, while another was placed extracardially and served as a reference lead. Forty‐six of the patients were followed postoperatively with measurements of myocardial stimulation threshold and resistance. In 25 of the patients, electrograms were recorded on magnetic tape for further computer analysis of amplitudes, slew rates, and signal source impedance. During constant current pacing, myocardial stimulation threshold increased from a median of 0.4 mA one hour postoperatively to a maximum value of 2.3 mA. In two patients (4.3%) intermittent pacing failure was seen. Stimulation resistance fell from a median of 875 Ω, to a minimum of 487Ω, with a subsequent increase to 598Ω before electrode removal. Both mean electrogram amplitude (7.35 mV) and slew rate (0.82 V/s) had their minimum values on the sixth postoperative day. Intermittent sensing failure was observed in 2/25 patients (8%). Signal source impedance was of a magnitude not likely to contribute to sensing failure. No complications were seen from the use of this lead. The new electrode is an important improvement in temporary pacemaker lead design. (PACE, Vol. 5, July‐August, 1982)Keywords
This publication has 11 references indexed in Scilit:
- R-wave amplitude variations during acute experimental myocardial ischemia: an inadequate index for changes in intracardiac volume.Circulation, 1981
- Signal source impedance of implanted pacemaker electrodes estimated from the spectral ratio between loaded and unloaded electrograms in manMedical & Biological Engineering & Computing, 1980
- Temporary Pacemaker Treatment in Open Heart Surgery: Pre‐ to Postoperative Changes in the Electrogram CharacteristicsPacing and Clinical Electrophysiology, 1980
- The Interdependence Between Electrogram, Total Electrode Impedance and Pacemaker Input Impedance Necessary to Obtain Adequate Functioning of Demand PacemakersPacing and Clinical Electrophysiology, 1979
- Temporary Pacemaker Treatment in Open Heart Surgery: Variation in Myocardial Threshold, Tissue and Interface Impedances in Man*Pacing and Clinical Electrophysiology, 1979
- Pronounced dependence of ventricular endocardial QRS potentials on ventricular volume.Heart, 1978
- Cardiac pacing and pacemakers III. Sensing the cardiac electrogramAmerican Heart Journal, 1977
- An improved implantable temporary pacemaker electrodeThe Journal of Thoracic and Cardiovascular Surgery, 1974
- Temporary Postoperative Epicardial Pacing ElectrodesThe Annals of Thoracic Surgery, 1969
- Cardiac Pacemaking in the Management of Postoperative ArrhythmiasThe Annals of Thoracic Surgery, 1968