ON THE OPTICAL MEASUREMENT OF CORNEAL THICKNESS
Authors:
- 1 October 1980
- journal article
- research article
- Published by Wiley in Acta Ophthalmologica
- Vol. 58 (5) , 760-766
Abstract
The optical principle for the estimation of corneal thickness with a commonly employed method is considered. The theoretical relationship between the apparent and true corneal thickness is described in a simple manner. Based on the theoretical results it could be confirmed that physiological variations in refractive index and radius of the cornea induce an insignificant error in the corneal thickness estimate. The theoretical relationship between apparent and true corneal thickness was compared to the performance of two commercially available pachometers (Haag‐Streit and Zeiss). The actual reading of the pachometers was found to be lower than the theoretical reading by a non‐linear error amounting to about 0.010 and 0.100 mm at a true corneal thickness of 0.50 and 1.00 mm, respectively.Keywords
This publication has 4 references indexed in Scilit:
- A TECHNICAL IMPROVEMENT OF THE HAAG‐STREIT PACHOMETER Short CommunicationActa Ophthalmologica, 1977
- ON THE OPTICAL MEASUREMENT OF CORNEAL THICKNESSActa Ophthalmologica, 1971
- Measurement of Corneal Thickness With the Haag-Streit PachometerArchives of Ophthalmology (1950), 1968
- Tiefenmessung der menschlichen Vorderkammer mit planparallelen Platten (Zusatzger t zur Spaltlampe)Graefe's Archive for Clinical and Experimental Ophthalmology, 1952