Global robotic experience and the type of surgical system impact the types of robotic malfunctions and their clinical consequences: an FDA MAUDE review
- 1 November 2011
- journal article
- research article
- Published by Wiley in BJU International
- Vol. 109 (8) , 1222-1227
Abstract
Study Type – Therapy (case series) Level of Evidence 4 What's known on the subject? and What does the study add? Malfunctions of the robotic surgical system have been reported, and the critical failure rate leading to converting or aborting the case occurs in less than 1% of cases. However, little is known about how global robotic experience with time and the advent of newer robotic surgical systems impact robot malfunctions. In this study, we characterize the changes in type and consequences of robot malfunctions over time and by type of robotic system used (da Vinci or da Vinci S). OBJECTIVES: To assess annual rates of robotic system malfunctions and compare the da Vinci S® system (dVS) and da Vinci® surgical system (dV). To assess the types of malfunctions and associated outcomes for robotic cases and determine the extent to which experience and technological improvements impact these. PATIENTS AND METHODS: This study is a retrospective review of the US Food and Drug Administration (FDA) MAUDE (Manufacturer and User Facility Device Experience) database, a publicly available, voluntary reporting system ( http://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/search.cfm ). The database was searched using the two phrases ‘da Vinci’ and ‘Intuitive Surgical’ from 2003 to 2009. Malfunctions of the instruments, console, patient‐side cart, camera and cannula were recorded. Data on intraoperative injuries, case delays and conversions were also collected. RESULTS: In all, 1914 reports were reviewed (991 dVS and 878 dV, 45 unclassified) with peak years for reports of 2008 for dVS (571) and 2007 for dV (211), P < 0.001. With respect to time, the proportion of console and patient‐side cart malfunctions declined from 2007 onward compared with the proportions prior to 2007 (5.1% vs 9.4% and 6.6% vs 10.9%). Patient injury did not change with year of surgery (0.5–5.4% of malfunctions, P = 0.358), open conversions declined (21.3% of malfunctions before 2007 vs 9.9% from 2007 onward, P < 0.001) and patient deaths increased (0.0013% of cases before 2007 vs 0.0061% of cases from 2007 onward, P < 0.001). With regard to robotic system, console and patient‐side cart malfunctions were more frequent with the dV than the dVS: 82/878 vs 39/991 and 100/878 vs 48/991, P < 0.001. Open conversion was more frequent with dV than dVS (19.3% vs 7.7% of reported malfunctions, P < 0.001), while patient injury was less with dV than dVS (3.5% vs 5.9%, P = 0.021). CONCLUSIONS: The dVS decreased console and patient‐side cart errors relative to total malfunctions, which were also influenced by surgical year. Open conversions were reduced by increased robotic experience and newer surgical system. Differences in patient injury may reflect changes in reporting or case complexity.Keywords
This publication has 9 references indexed in Scilit:
- The role of the U.S. Food and Drug Administration in device evaluation and monitoringGastrointestinal Endoscopy, 2010
- Critical appraisal of technical problems with robotic urological surgeryBJU International, 2010
- Failure and Malfunction of da Vinci Surgical Systems During Various Robotic Surgeries: Experience From Six Departments at a Single InstituteUrology, 2009
- Robotic Equipment Malfunction During Robotic Prostatectomy: A Multi-institutional StudyJournal of Endourology, 2008
- Introduction to the MAUDE DatabaseClinical Obstetrics & Gynecology, 2008
- Intraoperative Breakage of Needle Driver Jaw During Robotic-Assisted Laparoscopic Radical ProstatectomyUrology, 2008
- Da Vinci Robot Error and Failure Rates: Single Institution Experience on a Single Three-Arm Robot Unit of More than 700 Consecutive Robot-Assisted Laparoscopic Radical ProstatectomiesJournal of Endourology, 2007
- Robotic radical prostatectomy and the vattikuti urology institute technique: an interim analysis of results and technical pointsUrology, 2003
- An intraoperative method of localizing a missing piece of a broken laparoscopic instrumentAmerican Journal of Obstetrics and Gynecology, 1997