Incidence of Atrial Fibrillation in Patients with an Insertable Cardiac Monitor and Symptomatic Heart Failure
Open Access
- 9 December 2024
- journal article
- research article
- Published by Oxford University Press (OUP) in ESC Heart Failure
- Vol. 12 (3) , 1693-1702
Abstract
Aims: We aim to evaluate the incidence of atrial fibrillation (AF) in a large real-world cohort of patients implanted with an insertable cardiac monitor (ICM) who had a clinical history of symptomatic heart failure (HF) with reduced or preserved left ventricular ejection fraction (LVEF). Methods: Patients with an ICM and a history of HF events were identified from the Optum® de-identified Electronic Health Record dataset merged with an ICM device dataset collected during 2007–2021. All ICM-detected AF episodes that were available with ≥30-s of ECG at onset were adjudicated using artificial intelligence (AI model). Episodes with AI model probability of AF ≥ 0.9 were analysed. The Kaplan–Meier incidence of AF as a function of episode duration, history of AF, and LVEF were assessed. Results: A total of 1020 patients with ICM were identified of whom 911 had ≥180 days of follow-up and were included. According to the AI model, 358 patients had 8407 episodes of true AF. Incidence of AF at 42 months was 45.6% (44.1% vs. 46.8% in reduced vs. preserved LVEF). Incidence of new-onset AF was 23.2% (23.3% vs. 22.2% in reduced vs. preserved LVEF) in patients with no clinical history of AF. Patients with new-onset AF had a higher HF event rate compared with patients who had no clinical history of AF and did not develop AF during follow-up [OR = 2.73 (1.47–5.09); P = 0.002]. Patients with preserved LVEF had more longer duration paroxysmal AF compared with those with reduced LVEF (44.5% vs. 33.9%, P = 0.02). Conclusions: AF was observed in almost half of patients with ICM and symptomatic HF. One-fourth of the patients had new onset AF and a higher rate of HF events compared with patients without AF. AF incidence was similar in patients with preserved and reduced LVEF.Keywords
Funding Information
- Medtronic
- Medtronic
This publication has 33 references indexed in Scilit:
- Detection of subclinical atrial fibrillation with cardiac implanted electronic devices: What decision making on anticoagulation after the NOAH and ARTESiA trials?European Journal of Internal Medicine, 2024
- Atrial Fibrillation In Patients With Stroke Attributed to Large- or Small-Vessel DiseaseJAMA Neurology, 2023
- An ensemble of features based deep learning neural network for reduction of inappropriate atrial fibrillation detection in implantable cardiac monitorsHeart Rhythm O2, 2022
- Real-World Incidence of Pacemaker and Defibrillator Implantation Following Diagnostic Monitoring With an Insertable Cardiac MonitorThe American Journal of Cardiology, 2019
- Incidence of Previously Undiagnosed Atrial Fibrillation Using Insertable Cardiac Monitors in a High-Risk PopulationJAMA Cardiology, 2017
- Silent atrial fibrillation: epidemiology, diagnosis, and clinical impactClinical Cardiology, 2017
- Heart Failure With Preserved Ejection Fraction and Atrial FibrillationJACC, 2016
- Real-world performance of an enhanced atrial fibrillation detection algorithm in an insertable cardiac monitorHeart Rhythm, 2016
- Discerning the Incidence of Symptomatic and Asymptomatic Episodes of Atrial Fibrillation Before and After Catheter Ablation (DISCERN AF)JAMA Internal Medicine, 2013
- Detection of asymptomatic arrhythmias in unexplained syncopeAmerican Heart Journal, 2004