The Potential Risk of Conical Implant‐Abutment Connections: The Antirotational Ability of Cowell Implant System
Authors:
- 28 March 2014
- journal article
- research article
- Published by Wiley in Clinical Implant Dentistry and Related Research
- Vol. 17 (6) , 1208-1216
Abstract
Background: Conical implant‐abutment connections are popular because of good antibending performance. However, the cross section is round, and the antirotational ability is questionable because restorations in the oral cavity also have to bear torsional moments resulting from chewing patterns.Purpose: The purpose of this study was to investigate the antirotational ability of conical implant‐abutment connections with and without an index.Materials and Methods: Conical connection implant system (Cowell Medi, Busan, South Korea) was selected. Two kinds of cyclic loading, a bending moment with (C) and without (L) a torsional moment, were respectively applied to two kinds of abutments, pure cone (N‐Octa) and cone with an octagonal index (Octa). The number of cycles to fatigue and the failure modes was recorded. Morphologies of the abutments were examined with scanning electron microscopy.Results: Only group C(N‐Octa) passed the fatigue test, whereas the other three groups failed because of different failure modes. In group L(N‐Octa), all abutments generated rotation within 150 cycles. In groups C(Octa) and L(Octa), all abutments fractured but in different areas.Conclusions: In Cowell implant system (taper angle = 7°), there was no antirotational ability in purely conical connections. Adding an octagonal index could provide an antirotational function but could compromise the antibending strength of the abutment.Keywords
Funding Information
- National Science Council (101-2314-B-010-035)
This publication has 20 references indexed in Scilit:
- Effects of different implant–abutment connections on micromotion and stress distribution: Prediction of microgap formationJournal of Dentistry, 2012
- The effect of different implant-abutment connection on screw joint stability.Journal of Oral Implantology, 2012
- Influence of microgap location and configuration on the periimplant bone morphology in submerged implants. An experimental study in dogsClinical Oral Implants Research, 2008
- A systematic review of the 5‐year survival and complication rates of implant‐supported single crownsClinical Oral Implants Research, 2007
- In vitro differences of stress concentrations for internal and external hex implant–abutment connections: a short communicationJournal of Oral Rehabilitation, 2006
- Dynamic fatigue resistance of implant–abutment junction in an internally notched morse‐taper oral implant: influence of abutment designClinical Oral Implants Research, 2004
- Role of the Microgap Between Implant and Abutment: A Retrospective Histologic Evaluation in MonkeysJournal of Periodontology, 2003
- Influence of the Size of the Microgap on Crestal Bone Changes Around Titanium Implants. A Histometric Evaluation of Unloaded Non‐Submerged Implants in the Canine MandibleJournal of Periodontology, 2001
- The development of the ITI® DENTAL IMPLANT SYSTEMClinical Oral Implants Research, 2000
- The development of the ITI® DENTAL IMPLANT SYSTEMClinical Oral Implants Research, 2000