Comparison of the automated non-invasive oscillometric blood pressure monitor (BpTRU???) with the auscultatory mercury sphygmomanometer in a paediatric population
Authors:
- Balraj S. HeranBHBalraj S. Heran
- James M. WrightJWJames M. Wright
- 1 February 2004
- journal article
- research article
- Published by Wolters Kluwer Health in Blood Pressure Monitoring
- Vol. 9 (1) , 39-45
Abstract
To directly compare the accuracy of the BpTRU™ (an automated oscillometric blood pressure device) with standard auscultatory mercury sphygmomanometry in a pediatric population. The BpTRU™ was connected in parallel with a standard mercury sphygmomanometer. Two observers measured the blood pressures at the same time as it was being measured by the BpTRU™. The observers and the BpTRU™ were all blinded from each other. For each of the demographic data—subject age, sex and arm sizes—the mean, standard deviation (SD) and range was calculated. The difference between the mean BpTRU™ and the standard reference measurements (observer average) was calculated with SD and ranges. The percentage of measurements within 5, 10 and 15 mmHg agreement was expressed. From the 36 subjects recruited aged 3–18 years, 162 pairs of sitting blood pressures were included. The difference between the mean BpTRU™ readings and the reference standard measurements (as determined by the observers) was 1.45±5.67 mmHg for systolic blood pressures, and –3.24±7.39 mmHg for diastolic pressure and 0.20±2.47 bpm for heart rate. The BpTRU™ is of similar accuracy in measuring blood pressure in children as it was in an adult population.Keywords
This publication has 5 references indexed in Scilit:
- Pediatric hypertension: recent literatureCurrent Opinion in Pediatrics, 2002
- How should we measure blood pressure in the doctorʼs office?Blood Pressure Monitoring, 2001
- Comparison of the oscillometric blood pressure monitor (BPM-100 Beta ) with the auscultatory mercury sphygmomanometerBlood Pressure Monitoring, 2001
- Validation of a new algorithm for the BPM-100 electronic oscillometric office blood pressure monitorBlood Pressure Monitoring, 2001
- Silver-Russell Syndrome With Unusual FindingsPediatrics, 1987