Design and Characterisation of a Polyvinyl Chloride (PVC) Tissue-Mimicking Polymer Phantom for Quantitative Shear Wave Elastography Validation
Open Access
Abstract
A polyvinyl chloride (PVC)-based tissue-mimicking polymer phantom was developed and mechanically characterised to replicate stiffness ranges relevant to breast elastography and to provide a controlled platform for evaluating shear wave elastography (SWE) measurements. SWE provides quantitative stiffness information that complements B-mode ultrasound in breast imaging. However, measurement variability related to operator technique and tissue continues to limit confidence in clinical interpretation. This study evaluates the reproducibility of SWE using custom-fabricated PVC-based breast phantoms with mechanically defined stiffness properties. Two PVC-based breast phantoms with identical geometry and different background stiffnesses were scanned using a single ultrasound system under a fixed SWE protocol. Each phantom contained four embedded inclusions representing clinically relevant stiffness categories. Six breast imagers independently acquired repeated SWE measurements in transverse and longitudinal planes, blinded to lesion identity and ground truth. Inter-operator reproducibility was assessed using intraclass correlation coefficients, and was high across both phantom backgrounds, with low intra-operator variability following quality assurance exclusion of one dataset due to sampling error. Measurement variability was lowest for solid inclusions and increased for the cyst-like inclusion in the stiffer background. SWE measurements consistently preserved the relative stiffness ordering of inclusions, although absolute values differed systematically from mechanically derived ground-truth stiffness. These findings demonstrate that PVC-based polymer phantoms provide a stable and reproducible platform for evaluating SWE measurement behaviour under controlled conditions. By isolating operator and acquisition effects from biological variability, this polymer-based framework supports methodological standardisation and structured operator training in breast elastography.Keywords
This publication has 40 references indexed in Scilit:
- Intra‐ and Interobserver Reliability of Shear Wave Elastography in Breast Cancer DiagnosisJournal of Ultrasound in Medicine, 2023
- Diagnostic Accuracy of Four Levels of Manual Compression Applied in Supersonic Shear Wave Elastography of the BreastAcademic Radiology, 2021
- Individualized-thresholding Shear Wave Elastography combined with clinical factors improves specificity in discriminating breast massesThe Breast, 2020
- Audit of eliminating biopsy for presumed fibroadenomas with benign ultrasound greyscale and shear-wave elastography findings in women aged 25–39 yearsClinical Radiology, 2020
- Comparison of five viscoelastic models for estimating viscoelastic parameters using ultrasound shear wave elastographyJournal of the Mechanical Behavior of Biomedical Materials, 2018
- Stiffness at shear-wave elastography and patient presentation predicts upgrade at surgery following an ultrasound-guided core biopsy diagnosis of ductal carcinoma in situClinical Radiology, 2016
- Validation of intra- and interobserver reproducibility of shearwave elastography: Phantom studyUltrasonics, 2013
- Ultrasound elastography: Principles and techniquesDiagnostic and Interventional Imaging, 2013
- Breast elastography: A literature reviewJournal of Ultrasound, 2012
- Shear wave elastography for breast masses is highly reproducibleEuropean Radiology, 2011