Ambient dose during intraoral radiography with current techniques: part 3: effect of tube voltage
- 1 May 2021
- journal article
- research article
- Published by Oxford University Press (OUP) in Dentomaxillofacial Radiology
- Vol. 50 (4) , 20190362
Abstract
This study was conducted to assess the impact of tube voltage on ambient dose during intraoral radiography, specifically remnant-beam transmission and scattered radiation. Remnant-beam and scattered radiation doses were recorded using a phantom at tube voltages of 60, 63, 66, and 70 kV. Mathematical equations depicting their relations were then formulated, and reference values were calculated at the various voltages tested. Total ambient doses per exposure at 60 kV and at 70 kV were compared. Both remnant-beam transmission and scattered radiation increased ~40% by increasing tube voltage from 60 kV to 70 kV, and the relation was linear. A remnant beam transmission reference value of 7.5% was established at 70 kV, as well as a conversion factor of 0.035µSv/mAs at one meter for scattered radiation at 60 kV. Given longer exposure times at 60 kV, total ambient dose proved higher at 60 kV than at 70 kV. Higher tube voltage results in higher remnant-beam transmission and more scattered radiation per workload. The relation is linear in the range between 60kV and 70 kV. Remnant-beam transmission at 70 kV is safely assessed at 7.5%. A conversion factor at 60 kV of 0.035 µSv/mAs at one meter for the scattered radiation dose can be proposed. Total ambient dose per exposure was higher at 60kV than at 70 kV.Keywords
This publication has 9 references indexed in Scilit:
- Digital dental radiology in Belgium: a nationwide surveyDentomaxillofacial Radiology, 2018
- Ambient dose during intra-oral radiography with current techniques: Part 2 quantifying the remnant beam - an in vivo studyDentomaxillofacial Radiology, 2018
- Ambient dose during intraoral radiography with current techniques: Part 1 conversion factor for scattered radiation using a rectangular collimatorDentomaxillofacial Radiology, 2018
- Evaluation of mean conversion coefficients from air-kerma toH*(10) using secondary and transmitted x-ray spectra in the diagnostic radiology energy rangeJournal of Radiological Protection, 2016
- Distribution of scattered radiation during intraoral radiography with the patient in supine positionOral Surgery Oral Medicine Oral Pathology Oral Radiology and Endodontology, 1998
- IntroductionAnnals of the ICRP, 1991
- Effective dose equivalent to the operator in intra-oral dental radiography.Dentomaxillofacial Radiology, 1990
- Radiation isoexposure curves about a dental chair during radiographyThe Journal of the American Dental Association, 1969
- Secondary radiation and the dentistThe Journal of the American Dental Association, 1958