An evaluation of the N-CAT, a new arterial tonometer
Authors:
Open Access
- 1 June 1995
- journal article
- equipment
- Published by Springer Nature in Canadian Journal of Anesthesia/Journal canadien d'anesthésie
- Vol. 42 (6) , 526-531
Abstract
The N-CAT is a newly developed arterial tonometer (TBP) monitor able to determine systolic, diastolic and mean blood pressures continuously and noninvasively. The aim of this study was to evaluate the accuracy and reliability of the TBP compared with directly measured invasive blood pressure (IBP) in 14 patients before and after elective coronary artery bypass surgery. Although the TBP was able to track changes in systemic pressure, before and after CPB, bias and precision for TBP monitoring did not meet the standard criteria for equivalency for noninvasive blood pressure to invasive blood pressure. We were unable to monitor TBP in two patients. Approximately 40% of all before and after CPB mean TBP pressure values differed from mean IBP by more than 10 mmHg. Moreover, there were discrepancies of sufficient magnitude and duration that limits the clinical usefulness of the N- CAT. Potential users should not rely exclusively on TBP values when making clinical decisions. Technological improvement is needed before its clinical use is recommended. Le N-CAT est un nouveau moniteur de pression artérielle qui utilise le principe de la tonométrie artérielle (TBP) pour déterminer la pression systolique, diastolique, et moyenne de façon continue et non invasive. Le but de cette étude est d’évaluer la précision et la fiabilité de la TBP par rapport à la pression artérielle invasive (IBP) chez des patients avant une chirurgie cardiaque (pré- CEC) et après chirurgie cardiaque (post- CEC). La TBP a été capable de détecter les changements de pression. Cependant, les valeurs de biais et de précision pré- et post-CEC n’ont pas rencontré les critères de l’Association for the Advancement of Medical Instrumentation. Nous avons également été incapables de monitorer la TBP chez deux patients. Environ 40,3% des différences entre la TBP et l’IBP moyenne se situaient dans l’intervalle supérieur à 10 mmHg pré- CEC et 41,2% post- CEC. De plus, il y a eu des moments où la différence entre la TBP et l’IBP était de durée et de magnitude non acceptables, limitant son utilité chez des patients où la mesure exacte et continue de la pression artérielle est souhaitable. Des améliorations technologiques seraient souhaitables avant que son utilisation clinique soit recommandée.Keywords
This publication has 13 references indexed in Scilit:
- Comparison of Arterial Tonometry with Radial Artery Catheter Measurements of Blood Pressure in Anesthetized PatientsAnesthesiology, 1994
- Assessment of the arterial tonometer (N-CAT) for the continuous blood pressure measurement in rapid atrial fibrillationCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1993
- Arterial Tonometry for Noninvasive, Continuous Blood Pressure Monitoring during AnesthesiaAnesthesiology, 1991
- Blood Pressure Measurement by Arterial Tonometry in Controlled HypotensionAnesthesia & Analgesia, 1991
- The Accuracy of Finapres™ Noninvasive Mean Arterial Pressure Measurements in Anesthetized PatientsAnesthesiology, 1991
- Accuracy and dynamic response of disposable pressure transducertubing systemsCanadian Journal of Anesthesia/Journal canadien d'anesthésie, 1987
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986
- Can We Trust the Direct Radial Artery Pressure Immediately Following Cardiopulmonary Bypass?Anesthesiology, 1985
- Arterial tonometry: Review and analysisJournal of Biomechanics, 1983
- Direct Blood Pressure Measurement —Dynamic Response RequirementsAnesthesiology, 1981