In-vivo validation of spatially correct three-dimensional reconstruction of human coronary arteries by integrating intravascular ultrasound and biplane angiography
Authors:
- Yiannis S. ChatzizisisYCYiannis S. Chatzizisis
- George SianosGSGeorge Sianos
- Dimitrios TsikaderisDTDimitrios Tsikaderis
- Antonis MatakosAMAntonis Matakos
- Vassilios KoutkiasVKVassilios Koutkias
- Panagiotis DiamantopoulosPDPanagiotis Diamantopoulos
- Nicos MaglaverasNMNicos Maglaveras
- George E. ParcharidisGPGeorge E. Parcharidis
- George E. LouridasGLGeorge E. Louridas
- 1 September 2006
- journal article
- research article
- Published by Wolters Kluwer Health in Coronary Artery Disease
- Vol. 17 (6) , 533-543
Abstract
The in-vivo validation of geometrically correct three-dimensional reconstruction of human coronary arteries by integrating intravascular ultrasound and biplane coronary angiography has not been adequately investigated. The purpose of this study was to describe the reconstruction method and investigate its in-vivo feasibility and accuracy. In 17 coronary arteries (mean length, 85.7±17.1 mm) from nine patients, an intravascular ultrasound procedure along with a biplane coronary angiography was performed. From each angiographic projection, a single end-diastolic frame was selected in order to reconstruct the intravascular ultrasound catheter trajectory in space. In each end-diastolic intravascular ultrasound image, the lumen and media–adventitia contours were detected semi-automatically by an active contour algorithm. Each pair of contours was located on the catheter trajectory appropriately and interpolated with the adjacent pairs creating a three-dimensional volume of the arterial lumen and wall. The reconstructed lumen was back-projected onto both angiographic planes and the agreement between the back-projected and the angiographic luminal outlines was calculated. The angiogram-derived catheter length showed very high correlation (y=0.97x+1.8, P<0.001) and agreement with the corresponding pullback-derived values. Accordingly, the semi-automated segmentation of intravascular ultrasound images was also in significant correlation (r≥0.96, P<0.001) and agreement with the reference manual tracing. The back-projected luminal borders showed good overall association with the corresponding angiographic ones (r=0.78, P<0.001) as well as remarkable agreement. Spatially correct three-dimensional reconstruction of human coronary arteries constitutes an imaging method with considerably high in-vivo feasibility and accuracy.Keywords
This publication has 15 references indexed in Scilit:
- Reproducibility of coronary lumen, plaque, and vessel wall reconstruction and of endothelial shear stress measurements in vivo in humansCatheterization and Cardiovascular Interventions, 2003
- American College of Cardiology clinical expert consensus document on standards for acquisition, measurement and reporting of intravascular ultrasound studies (ivus)31When citing this document, the American College of Cardiology would appreciate the following citation format: Mintz GS, Nissen SE, Anderson WD, Bailey SR, Erbel R, Fitzgerald PJ, Pinto FJ, Rosenfield K, Siegel RJ, Tuzcu EM, Yock PG. ACC Clinical Expert Consensus Document on Standards for the acquisition, measurement and reporting of intravascular ultrasound studies: a report of the American College of Cardiology Task Force on Clinical Expert Consensus Documents (Committee to Develop a Clinical Expert Consensus Document on Standards for Acquisition, Measurement and Reporting of Intravascular Ultrasound Studies [IVUS]). J Am Coll Cardiol2001;37:1478–92.33Address for reprints: This document is available on the ACC Website at www.acc.org. Reprints of this document are available for $5.00 each by calling 800-253-4636 (U.S. only) or writing to the Resource Center, American College of Cardiology, Educational Services, 9111 Old Georgetown Road, Bethesda, MD 20814-1699.JACC, 2001
- True 3-Dimensional Reconstruction of Coronary Arteries in Patients by Fusion of Angiography and IVUS (ANGUS) and Its Quantitative ValidationCirculation, 2000
- Deformable boundary detection of stents in angiographic imagesIEEE Transactions on Medical Imaging, 2000
- Three-dimensional reconstruction of the coronary artery wall by image fusion of intravascular ultrasound and bi-plane angiographyThe International Journal of Cardiovascular Imaging, 2000
- Geometrically correct 3-D reconstruction of intravascular ultrasound images by fusion with biplane angiography-methods and validationIEEE Transactions on Medical Imaging, 1999
- An attempt to 3D reconstruct vessel morphology from X-ray projections and intravascular ultrasounds modeling and fusionComputerized Medical Imaging and Graphics, 1996
- Accurate Three-Dimensional Reconstruction of Intravascular Ultrasound DataCirculation, 1996
- Snakes: Active contour modelsInternational Journal of Computer Vision, 1988
- STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENTThe Lancet, 1986