A Method for Optimizing and Validating Institution-Specific Flagging Criteria for Automated Cell Counters
Authors:
- Anthony SireciASAnthony Sireci
- Robert SchlabergRSRobert Schlaberg
- Alexander KratzAKAlexander Kratz
Open Access
- 1 October 2010
- journal article
- Published by Archives of Pathology and Laboratory Medicine in Archives of Pathology & Laboratory Medicine
- Vol. 134 (10) , 1528-1533
Abstract
Context. —Automated cell counters use alerts (flags) to indicate which differential white blood cell counts can be released directly from the instrument and which samples require labor-intensive slide reviews. The thresholds at which many of these flags are triggered can be adjusted by individual laboratories. Many users, however, use factory-default settings or adjust the thresholds through a process of trial and error. Objective. —To develop a systematic method, combining statistical analysis and clinical judgment, to optimize the flagging thresholds on automated cell counters. Design. —Data from 502 samples flagged by Sysmex XE-2100/5000 (Sysmex, Kobe, Japan) instruments, with at least 1 of 5 user-adjustable, white blood cell count flags, were used to change the flagging thresholds for maximal diagnostic effectiveness by optimizing the Youden index for each flag (the optimization set). The optimized thresholds were then validated with a second set of 378 samples (the validation set). Results. —Use of the new thresholds reduced the review rate caused by the 5 flags from 6.5% to 2.9% and improved the positive predictive value of the flagging system for any abnormality from 27% to 37%. Conclusions. —This method can be used to optimize thresholds for flag alerts on automated cell counters of any type and to improve the overall positive predictive value of the flagging system at the expense of a reduction in the negative predictive value. A reduced manual review rate helps to focus resources on differential white blood cell counts that are of clinical significance and may improve turnaround time.Keywords
This publication has 13 references indexed in Scilit:
- The divergent morphological classification of variant lymphocytes in blood smearsJournal of Clinical Pathology, 2007
- Performance Evaluation of the CellaVision DM96 SystemAmerican Journal of Clinical Pathology, 2005
- The International Consensus Group for Hematology Review: Suggested Criteria for Action Following Automated CBC and WBC Differential AnalysisLaboratory Hematology, 2005
- Optimal Cut-point and Its Corresponding Youden Index to Discriminate Individuals Using Pooled Blood SamplesEpidemiology, 2005
- Immature Granulocyte Measurement Using the Sysmex XE-2100American Journal of Clinical Pathology, 2003
- Clinical utility of the band countClinics In Laboratory Medicine, 2002
- Peripheral blood film review: the demise of the eyecount leukocyte differentialClinics In Laboratory Medicine, 2002
- REGRET GRAPHS, DIAGNOSTIC UNCERTAINTY AND YOUDEN'S INDEXStatistics in Medicine, 1996
- Evaluation of the Leukocyte Differential Flags on an Hematologic Analyzer:The Cobas Argos 5 Diff.American Journal of Clinical Pathology, 1995
- Leukocyte DifferentialMayo Clinic Proceedings, 1993