Determination of dithiocarbamates; improvements to the method described in EN 12396-1 standard
- 6 February 2009
- journal article
- research article
- Published by Taylor & Francis in International Journal of Environmental Analytical Chemistry
- Vol. 89 (2) , 83-93
Abstract
A procedure to improve the performance and reliability of the method described in European standard EN-12396-1 is presented. A CS2 working range of 10–50 µg per 25 mL was validated. The quantification limit was lowered from 0.25 mg kg−1 to 0.05 mg kg−1 (200 g of sample matrix), in agreement with the Maximum Residue Limits for dithiocarbamates (DTCs). The procedure involves the measurement of the average absorbance at 374 nm and 435 nm, to allow for the equilibrium between the two coloured Cupric(II) complexes formed. Two main modifications of the decomposition and distillation apparatus assembly were introduced. One was the stabilization of the colour-forming reagent solution at 0°C to ensure stability of absorbance. The other was the inclusion of a bubbler trap with ethanol, to retain interfering compounds. The possible CS2 contamination by several types of protective gloves was also studied. Vinyl gloves, made according to the AQL 1.5 – EN 455-1/2 standard, did not interfere with the method.Keywords
This publication has 8 references indexed in Scilit:
- Environmental and pharmaceutical analysis of dithiocarbamatesAnalytica Chimica Acta, 2007
- Dithiocarbamates residues in Brazilian food and the potential risk for consumersFood and Chemical Toxicology, 2004
- Simultaneous determination of maneb and its main metabolites in tomatoes by liquid chromatography using diode array ultraviolet absorbance detectionJournal of Chromatography A, 2004
- Determination of dithiocarbamates and metabolites in plants by liquid chromatography–mass spectrometryJournal of Chromatography A, 2004
- Liquid Chromatography of Unstable Zinc Dithiocarbamates. Application to Rubber Gloves AnalysisJournal of High Resolution Chromatography, 2000
- Reasons for the decomposition of the fungicide thiram during preparation of fruit and vegetable samples and consequences for residue analysisAnalytical and Bioanalytical Chemistry, 2000
- Reduction of electronic noise in inductively coupled plasma atomic emission and fluorescence spectrometric measurementsAnalytical Chemistry, 1983
- Spectrophotometric Determination of Dithiocarbamate Residues in Food Crops.Analytical Chemistry, 1964