Human magnetic resonance imaging at 8 T
Authors:
- 1 October 1998
- journal article
- research article
- Published by Wiley in NMR in Biomedicine
- Vol. 11 (6) , 263-265
Abstract
In this work, we present the first human magnetic resonance image (MRI) obtained at ultrahigh field strengths (8 T). We demonstrate that clinical imaging will be possible at 8 T and that reasonable quality head images can be obtained at this field strength. Most importantly, we emphasize that the power required to excite the spins at 8 T is much lower than had previously been predicted by the nuclear magnetic resonance theory. A 90° pulse in the head at 8 T requires only ∼0.085 J of energy (90 W for a 2‐lobe 4 ms sinc pulse). Based on measurements at 4 T, 1–2 J of energy should have been utilized to achieve a 90° excitation at 8 T. The fact that the energy required for spin excitation at 8 T is much lower than predicted by the NMR theory, will be extremely important to the viability of ultrahigh field imaging, since concerns related to power absorption and specific absorption rate (SAR) violations at ultrahigh field are alleviated. As such, it will be possible to utilize RF intensive pulse sequences and adiabatic spin excitation at 8 T without significant risk to the subject. © 1998 John Wiley & Sons, Ltd.Keywords
This publication has 7 references indexed in Scilit:
- The design and test of a new volume coil for high field imagingMagnetic Resonance in Medicine, 1994
- High frequency volume coils for clinical NMR imaging and spectroscopyMagnetic Resonance in Medicine, 1994
- Spectroscopy and imaging with a 4 tesla whole‐body mr systemNMR in Biomedicine, 1988
- Image Formation by Induced Local Interactions: Examples Employing Nuclear Magnetic ResonanceNature, 1973
- XI. On the partition of energy between matter and ÆtherJournal of Computers in Education, 1905
- Ueber das Gesetz der Energieverteilung im NormalspectrumAnnalen der Physik, 1901
- LIII. Remarks upon the law of complete radiationJournal of Computers in Education, 1900